Analog/Digital; Physical Specifications - GE ApexPro TT Technical Manual

Designed to acquire and monitor physiological data for ambulating adult and pediatric patients within a defined coverage area.
Table of Contents

Advertisement

Technical specifications

Analog/digital

Physical specifications

A-4
Graph request:
Event Marker:
Dynamic range:
Input offset:
Input impedance:
ECG gain selection:
Gain accuracy:
Common mode rejection:
Defibrillator protection:
Defibrillation recovery time:
Pacemaker detection:
Power on/off:
Lead fail detection:
Serial communications:
A/D converter resolution:
Sample rate:
Height:
Width:
Depth:
Weight:
ApexPro™
Transmitted
Transmitted (ApexPro CH model only)
± 5 mV (RTI)
± 300 mV (RTI)
15 Megohm min differential at 10 Hz
5, 10, 20, 40 mm/mV (RTI)
± 5% at 15 Hz
100 dB min at 60 Hz
± 5000 VDC, 360 joules into 100 ohm load
Defibrillation recovery time: Limited only by electrode
recovery time. Transmitter recovers within 5 sec.
ApexPro model:
± 2 mV to ± 700 mV (RTI); 100 µsec to 2 msec; either
polarity
ApexPro CH model:
± 2 mV to ± 700 mV (RTI); 100 µsec to 2 msec; either
polarity
Multi-vector: Detection on 2 separate ECG vectors
(leads LL and Va)
Battery insertion/removal
DC type; indicates leadwire failed (i.e., RA, LA, LL, Va, or
Vb)
2 – 9600 baud asynchronous
(Note: Does not apply to units configured with inactive
interface connector ports.)
10 bits, 9.76 µV (RTI)
240 samples/sec
13.7 cm (5.4 in)
7.3 cm (2.9 in)
2.3 cm (0.9 in)
141.8 g (5 oz) Weight excludes batteries and leadwire
assembly.
2001989-351A