Environmental Specifications; Device Specifications - GE ApexPro TT Technical Manual

Designed to acquire and monitor physiological data for ambulating adult and pediatric patients within a defined coverage area.
Table of Contents

Advertisement

Technical specifications
ECG

Environmental specifications

Operating conditions
Transport and storage conditions

Device specifications

A-6
Multi-channel configuration
(5- or 6- leadwire)
Leads analyzed
simultaneously
Single-channel (3-leadwire)
configuration
Heart rate detection
QRS detection range
Frequency response
A/D converter resolution
Sample rate
Temperature
Relative humidity
Temperature
Relative humidity
Pressure
Water resistance
Input configuration
Frequency range
Power output:
Bandwidth
Modulation
Serial I/O ports
Alarm pause
ApexPro™
I, II, III, Va, Vb, aVR, aVL, aVF
Four (I, II, III, V)
I, II or III, configurable
30 to 300 BPM
0.5 to 5 mV
up to 38Hz (-3 dB)
10 bits, 9.75 µV (RTI)
240 samples/seconds
5 to 40° C
15 to 95% (non-condensing)
-40 to 70° C
15 to 95% (non-condensing)
475 to 1200 hPa
IEC 60529 IPX7 rating (can survive inadvertent submersion)
3, 5 or 6 electrodes
1395.025 - 1399.975 MHz
5mW
(Meets FCC requirements for Part 95.)
11.9 KHz
GFSK
2
Transmitted and indicated via LED
2001989-351A

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents