4. If, after six attempts to light the flame, it still is not lit, the ignition sequence terminates and the
hydrogen is shut off.
5. From the Chroma Board screen you can press the F4 (Manual Ignitor On) soft key to request
ignition. This automatically turns on the hydrogen supply.
If an analyzer with an FID has the Flameout LED illuminated on the front panel, it can be ignited
either automatically or manually. When the Chroma Board screen has Auto highlighted, the ignitor
will continuously try to ignite the flame. When Auto Timeout is highlighted, the ignitor will attempt to
ignite the flame for the time specified by Timeout on the Chroma Board screen.

Flame Ionization Detector (FID)

The FID cell (see Figure 4-3) reduces the physical sample to an electrical signal which an
interpretation device such as a recorder can condition and plot.
4-3
*****
Chroma Board
Detector: FID
Flame: Lit
Ignitor: Off
Hardware Zero:
Software Zero:
-
Detector Reading: 1.98mV
Ignitor: Manual Auto Auto Timeout
Timeout: 0 seconds
FPD Temp Zone: 0
Exit
Graphics
F1
Figure 4-2. CHROMA BOARD SCREEN
Breather
Adapter
Thermocouple
Ignitor
Figure 4-3. FLAME IONIZATION DETECTOR
H2: None
5.10mV (
4.7%)
3.18mV
Manually
Zero
Baseline
F2
F3
Microdot
Connector
Polarizer
Jet
*****
Manual
Ignitor
On
F4
2000-OSM, F1