Stream Table - ABB PGC2000 Operation And Service Manual

Process gas chromatograph
Table of Contents

Advertisement

10. Cursor right one column and select the value for the desired function (method number, valve
number, Vista Basic program, etc.). Use the keyboard number keys to enter the number and
then press ENTER.
11. Add more lines by inserting lines and repeating steps 6 through 10 as needed.
12. When you have completed the Time Table, press the F1 (Exit) soft key.
13. On the next screen press the F2 (Exit And Update) soft key to save the Time Table changes, or
press the F1 (Escape) soft key to retain the original values.
14. Press the F1 soft key as necessary to return to the Background screen.

Stream Table

The Stream Table identifies the status of each stream. This identifies the active streams and the
method number and Vista BASIC program assigned to each stream. It also allows you to edit the
table. You access the Stream Table for printing the same way as other tables, but you access it for
viewing or editing differently from other tables.
To view and edit a Stream Table:
1. On the Background screen, press the F1 (Exit to Commands) soft key.
2. On the Commands screen, cursor, if necessary, to ANALYSIS CONTROL and then press the F3
(Streams) soft key.
3. On the Stream Assignment screen (see Figure 4-31) review the stream data and edit as
necessary. This table lists the active streams and the method number and Vista BASIC program
number assigned to each stream. When "00" is in the Vista BASIC column, no Vista BASIC
program is associated with that stream.
2000-OSM, F1
******
Stream Status
Method Basic
01
Active 01
02
Inactive 01
03
Inactive 01
04
Inactive 01
05
Inactive 01
06
Inactive 01
07
Inactive 01
08
Inactive 01
Exit
F1
F2
Figure 4-31. STREAM ASSIGNMENT SCREEN
Stream Assignment
00
00
00
00
00
00
00
00
More Below
F3
******
F4
4-32

Advertisement

Table of Contents
loading

Table of Contents