1. On the Background screen press the F1 (Exit to Commands) soft key.
2. On the Commands screen cursor to ANALYSIS CONTROL, if necessary, and then press the F2
(Request Analysis) soft key.
3. On the Request Analysis screen (see Figure 3-5) select the method number to be calibrated.
4. Place the cursor on "Cal" and press the F2 (Accept The Request) soft key.
5. Press the F1 (Escape) soft key to return to the Commands screen.
6. On the Commands screen press the F4 (Start/Stop Analysis) soft key.
7. On the Start/Stop Analysis screen press the F2 (Start Analyzer) soft key. The controller will run
the calibration when all other requests have been completed.
8. When the calibration run is complete, verify that the Commands screen is displayed and then
press the F4 (Start/Stop Analysis) soft key.
9. On the Start/Stop Analysis screen, cursor to "At End of Analysis?" and then press the F2 (Stop
Analyzer) soft key.
10. Press the F1 soft key as necessary to return to the Background screen.
11. On the Background screen press the F3 (Cur Report, Basic I/O & Options) soft key. The screen
will display the current report. The analyzer will run the calibration when it clears other requests
and the report will appear on the screen as it happens.
12. When you have completed the calibration test, press the F1 soft key to return to the Background
screen.
3-4
******
Request Analysis
Analysis Requests:
or calibration enter the method#,
otherwise enter the stream#: [1
None
Stream
PENDING:
00
Accept
Escape
The
Request
F1
F2
Figure 3-5. REQUEST ANALYSIS SCREEN
For a benchmark
]
Bench
Cal
00
00
Clear
All
Requests
F3
******
F4
2000-OSM, F1

Advertisement

Table of Contents
loading

Table of Contents