Guidance And Manufacturer´s Declaration - GE Voluson S6 Basic User Manual

Hide thumbs Also See for Voluson S6:

Advertisement

16.12.6 Internal Battery
Type
Note
16.13 Guidance and manufacturer´s declaration
The Voluson® S6/S8 is inteded for use in electromagnetic enviroment specified below. The customer or
the user of the Voluson® S6/S8 should assure that it is used in such an enviroment.
Emission test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/ flicker
emissions
IEC 61000-3-3
The Voluson® S6/S8 is inteded for use in electromagnetic environment specified below. The customer or
the user of the Voluson® S6/S8 should assure that it is used in such an environment.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voluson® S6/S8 Basic User Manual
5433669-100 Revision 4
CR2450
Engraved "+" marking shall be on top position.
Guidance and manufacturer´s declaration - electromagnetic emissions
Compliance
Group 1
Class B
Class A
Complies
Guidance and manufacturer´s declaration - electromagnetic emissions
IEC 60601 test
Compliance level
level
± 2, 4, 6kV contact
± 2, 4, 6kV contact
± 2, 4, 8kV air
± 2, 4, 8kV air
± 2kV for power
± 2kV for power
supply lines
supply lines
±1kV for input/
±1kV for input/
output lines
output lines
± 1kV differential
± 1kV differential
mode
mode
± 2kV common
± 2kV common
mode
mode
Technical Data / Information
Electromagnetic enviroment - guidance
The Voluson® S6/S8 uses RF energy only for its
internal function. Therefore, its RF emissions are
very low and are not likely to cause any
interference in nearby electronic equipment.
The Voluson® S6/S8 is suitable for use in all
establishments, including domestic
establishments and those directly connected to
the public low-voltage power supply network that
supplies buildings used for domestic purposes,
provided the following warning in heeded:
WARNING: The Voluson® S6/S8 is intended for
use by healthcare professionals only. The
Voluson® S6/S8 may cause radio interference or
may disrupt the operation of nearby equipment.
It may be necessary to take mitigation measures,
such as re-orienting or relocating the system or
shielding the location.
Electromagnetic environment- guidance
Floors should be wood,concrete, or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Mains power quality should be that of a
typical commercial or hospital
environment.
Mains power quality should be that of a
typical commercial or hospital
environment.
16-31