Measurements and Patient Worksheets (Reports)
11.2.4.6 Circumference Measurements
Note
Note
11-42
1.
Press the [Calc] key on the control panel.
2.
Select the study for the corresponding item. For example: select [Fetal Biometry].
3.
Select the desired measurement parameter. For example: select [HC].
If HC or AC have been selected, you can select the desired way to perform the measurement,
using the left rotary control. Select from 2DAreaPoints, 2DArea Trace, AreaEllipse.
4.
Using the trackball, position the cursor on the perimeter of the shape to be measured.
Press the right or left trackball key [Set] to fix the mark. A second cursor appears.
5.
Move the second cursor (to form an appropriate ellipse) and press [Set] again.
To re-adjust the start point, press the upper trackball key [Change] before completing the
measurement. This alternates the control from one cursor to the other.
6.
Adjust the width of the ellipse by means of the trackball and then press the right trackball
key [Set].