Filling The Dispensing System - Miele PLW 8616 Operating Instructions Manual

Laboratory washer
Hide thumbs Also See for PLW 8616:
Table of Contents

Advertisement

Dispensing liquid chemical agents
Filling the dis-
pensing system
48
Push the suction lance into the neck of the container and screw it
into place.
Place the container in the drawer.
The suction lance must be located on the lower side of the support.
Push the drawer into the machine.
Insert the service panel into the lower brackets of the machine and
press on it.
Lock the service panel.
Acknowledge the message
with the OK button.
Each time a container is filled or replaced, the
function must be carried out to remove any air from the dispensing
system.
The function can only be carried out if there is no load carrier in the
wash cabinet.
A dispensing system must be filled when:
- The corresponding container has been filled or replaced with a new
one
- The process chemicals have been changed
A dispensing system can only be vented when there is no load car-
rier in the wash chamber.
System
Open the
Machine functions
Open the
Open the
Fill dispensing system
Press the button for the corresponding dispensing system, e.g.
DOS 1
.
The dispensing system will be filled.
Fill container for dispensing system "x"
menu.
option.
option.
Fill dispensing system