The following message will be displayed:
Starting setup for the dispensing system ... Preparing the dispensing
system.
Confirm the message in the display with OK.
The dispensed quantity will then be measured.
The process steps will be displayed, e.g.
When the measurement is complete, the on-screen keyboard opens.
The value set previously is shown.
Fill up the 500–1000 ml measuring cylinder up to the original fill
level with the process chemical from the 100 ml measuring cylinder.
Take a reading of the volume needed to fill the cylinder up from the
100 ml measuring cylinder.
Enter this value into the on-screen keyboard and confirm with OK.
The measurement results for the conversion factors are displayed,
e.g.:
Conversion factor DOS 1:
old ... µl/Imp new ... µl/Imp
Conversion factor flow sensor 1:
old ... µl/Imp new ... µl/Imp
Accept new values?
Select OK to apply the new conversion factors to the controller and
 to retain the old conversion factors. The
menu will appear again in the display.
Contact the Customer Service Department for the
system
if the machine's dispensing systems are not fitted with suc-
tion lances.
System settings
Dispenser pump active
Set up dispensing system
Set up dispensing
.
69