Plumbing

92
 Health risk and risk of damage due to contaminated incoming
water.
The quality of the incoming water must correspond to the drinking
water specification of the country in which the machine is being
operated.
Please refer to the installation diagram supplied with the machine.
- If the water supply has a high iron content, there is a danger that
corrosion may occur on load items and the machine itself.
- If the chloride content of the water exceeds 100 mg/l, the risk of
corrosion to items being cleaned in the machine will be increased
further.
- In certain regions, the water composition may cause precipitates to
form, requiring the use of softened water in the steam condenser.