Problem
The detector does not enter alarm.
The detector intermittently enters alarm
without apparent reason.
The detector automatically shuts off.
The detector displays: IntErnAL cLock
hAS FAiLEed
64
Table 16. Troubleshooting Tips (cont.)
Possible Cause
Alarm setpoint(s) are set incorrectly
Alarm setpoint(s) set to zero
Detector is in calibration mode
Ambient gas levels are near alarm
setpoint or the sensor is exposed to
a puff of the target gas
Alarms set incorrectly
Missing or faulty sensor
Automatic shutdown feature
activated due to weak batteries
General fault
Solution
Reset alarm setpoints (see
Calibration and Setting Alarm
Setpoints)
Reset alarm setpoints (see
Calibration and Setting Alarm
Setpoints)
Complete the calibration procedure
Detector is operating normally. Use
caution in suspect areas. Check
maximum gas exposure reading.
Reset alarm setpoints (see
Calibration and Setting Alarm
Setpoints)
Replace sensor (see
Replacing a
Sensor or Sensor
Filter)
Replace batteries (see
the
Batteries)
Contact
BW Technologies
Replacing