GasAlertMicro
User Manual
Problem
The detector does not enter alarm.
The detector intermittently enters alarm
without apparent reason.
The detector automatically shuts off.
The detector displays: IntErnAL cLock
hAS FAiLEed
64
Table 16. Troubleshooting Tips (cont.)
Possible Cause
→
Alarm setpoint(s) are set incorrectly
→
Alarm setpoint(s) set to zero
→
Detector is in calibration mode
→
Ambient gas levels are near alarm
setpoint or the sensor is exposed to
a puff of the target gas
→
Alarms set incorrectly
→
Missing or faulty sensor
→
Automatic shutdown feature
activated due to weak batteries
→
General fault
Solution
→
Reset alarm setpoints (see
Calibration and Setting Alarm
Setpoints)
→
Reset alarm setpoints (see
Calibration and Setting Alarm
Setpoints)
→
Complete the calibration procedure
→
Detector is operating normally. Use
caution in suspect areas. Check
maximum gas exposure reading.
→
Reset alarm setpoints (see
Calibration and Setting Alarm
Setpoints)
→
Replace sensor (see
Replacing a
Sensor or Sensor
Filter)
→
Replace batteries (see
the
Batteries)
→
Contact
BW Technologies
Replacing