The alarm setpoints can be modified only in the
order they are presented in this table. To bypass
any alarm setpoint, press C to retain the current
value and automatically proceed to the next alarm
setpoint.
Alarms may be set at any point within the detection
range for the sensor. Refer to
the alarm setpoint to zero to disable a sensor.
Factory alarm setpoints may vary by region.
Setting the TWA Alarm Setpoint
The TWA alarm setpoint for H
after the auto span, regardless if the span was successful or
not.
8.
Use G or H to select a new TWA alarm set point
value for H
S and press C to confirm the new
2
value.
Note
Note
Specifications
or set
S gas immediately displays
2
Calibration and Setting Alarm Setpoints
Or
To bypass the STEL alarm setpoint, press only C
to retain the current value and proceed to the low
setpoint.
If the new alarm setpoint is not confirmed within 10
seconds, the detector displays Err and
automatically retains the previous alarm setpoint.
Setting the STEL Alarm Setpoint
The STEL alarm setpoint displays automatically after the
TWA alarm setpoint.
9.
Use G or H to select a new STEL alarm setpoint
value for H
S and then press C to confirm the new
2
value.
Or
GasAlertMicro
Note
47