Honeywell BW GasAlertMicro User Manual page 73

Table of Contents

Advertisement

Problem
The detector does not display normal
ambient gas reading after the activation
self-test.
The detector does not respond to
pushbuttons.
The detector does not accurately
measure gas.
Table 16. Troubleshooting Tips (cont.)
Possible Cause
Sensor not stabilized
Detector requires calibration
Target gas is present
Batteries are depleted
Detector is performing operations
that do not require user input
Detector requires calibration
Detector is colder/hotter than
ambient gas
Sensor filter is blocked
GasAlertMicro
Troubleshooting
Solution
Used sensor: wait 60 seconds
New sensor: wait 5 minutes
Calibrate detector (see
and Setting Alarm
Setpoints)
Detector is operating properly. Use
caution in suspect areas.
Replace batteries (see
the
Batteries)
Pushbutton operation restored
automatically when the operation
ends
Calibrate sensor (see
Calibration
and Setting Alarm
Setpoints)
Allow the detector to acquire
ambient temperature before use
Clean the sensor filter (see
Replacing a Sensor or Sensor
Filter)
Calibration
Replacing
63