Problem
The detector does not display normal
ambient gas reading after the activation
self-test.
The detector does not respond to
pushbuttons.
The detector does not accurately
measure gas.
Table 16. Troubleshooting Tips (cont.)
Possible Cause
→
Sensor not stabilized
→
Detector requires calibration
→
Target gas is present
→
Batteries are depleted
→
Detector is performing operations
that do not require user input
→
Detector requires calibration
→
Detector is colder/hotter than
ambient gas
→
Sensor filter is blocked
GasAlertMicro
Troubleshooting
Solution
→
Used sensor: wait 60 seconds
New sensor: wait 5 minutes
→
Calibrate detector (see
and Setting Alarm
Setpoints)
→
Detector is operating properly. Use
caution in suspect areas.
→
Replace batteries (see
the
Batteries)
→
Pushbutton operation restored
automatically when the operation
ends
→
Calibrate sensor (see
Calibration
and Setting Alarm
Setpoints)
→
Allow the detector to acquire
ambient temperature before use
→
Clean the sensor filter (see
Replacing a Sensor or Sensor
Filter)
Calibration
Replacing