2-Point Temperature Calibration (Items Required)
-
Mini-AT
-
A cold temperature source (ice bath)
-
A hot temperature source (heated water, not to exceed 170 degrees F.)
-
Temperature Reference (calibrated thermometer)
-
Computer (IBM Compatible)
-
Mini-AT Link (or MasterLink32) Software
-
I/O cable, p/n 40-1629
Calibrated
Thermometer
Ice Bath
2-Point Temperature Calibration (Procedure)
Refer to figure 21 when temperature calibrating the Mini-AT. Since the two-point calibration is mostly a software
function, references to the appropriate sections in the Mini-AT Link software are included.
1) Prepare an agitated ice bath and a container of heated water.
2) Using serial cable p/n 40-1629, connect from the computer's serial port to the Mini-AT's RS-232
port.
www.honeywell.com
68
Page 68
Mini-AT User Guide
Pressure Transducer
Case Connector
Temperature
Probe
Calibrated
Thermometer
Heated
Water
2-Point Temperature Calibration
Mini-AT
Mini-AT
Link
Fig. 21
Back View
RS-232 Port
I/O Cable
p/n 40-1629