60
The 16-bit hexadecimal mask defined below can be specified
using the AE command. It is used for automatic error reporting
by the RE message transmitted by the instrument.
Bit Error
0
syntax error
1
parameter error
2
configuration
error
3
address error
4
checksum
error
5
zero error
6
calibration
error
7
sequence error
8
command not
available
9
range error
10
reserved
11
reserved
12
reserved
13
reserved
14
reserved
15
reserved
K0200 Issue No. 1
Table 6-1
Table of Error Definitions
Description
Set if the command syntax was not understood.
Set if the parameters in the command were out of
range or not valid.
The configuration parameters can only be changed
if the configuration PIN is sent with the commands. If
the PIN is not sent or is incorrect this error is reported.
An invalid address packet was received.
The received command checksum did not match
with the calculated checksum. With this error, the
command is not executed and this error reported.
An error occurred while trying to zero a measurand,
probably because the zero offset is too large.
Error in the instrument's calibration procedure;
not enough calibration points.
A valid command was received, but can not be
processed because the instrument was not in the
correct state to execute command.
This command was not available on this instrument.
Reading outside range.