Processing Measurements (Figure 3-6)
The instrument can process the measurement results in the
following ways:

Tare

Each measurement can be individually "tared". The tare
function substracts the presently measured value from all
subsequent measurements. When the tare function is
enabled, the display flashes the tare symbol. To enable and
disable the tare function, proceed as follows:
1.
Press the MODE push-button to select the measurement
to be tared.
2.
Press F1 to enter the instrument menu.
3.
Press F1 to view the process options.
4.
Press F1 again to view the tare options.
5.
Press F1 to enable tare or F2 to disable tare.

Filter

When switched on the low pass filter is applied to all
measurements in all three measurement modes. Before
switching on, the two parameters of BAND and TIME should
be set.
Note:When the time constant in the sample is large and the
filter is active at power-up, enough time must be allowed for
the sample to be received and a stable reading to be displayed.