GE Druck DPI 740 User Manual page 44

Precision pressure indicator
Table of Contents

Advertisement

36
Two-point calibration
Proceed as follows:
1.
Make sure the instrument is in measurement mode and
displaying pressure.
2.
Press SET until the display shows CAL on the status line.
3.
Press F1 to select the CAL menu.
4.
Use F1, F2 and SET to enter the PIN.
5.
Press F2 to select a two point calibration, the display
prompts for the first calibration point.
6.
Adjust the pressure standard for the first calibration point.
7.
Use F1
and F2
8.
When the pressure is stable press SET to enter the first
calibration point. The display prompts for the second
calibration point.
9.
Repeat steps 5 to 7 for the second calibration point.
10. Press F1 to accept the calibration values, the display
prompts for the calibration date.
11. Use F1
K0200 Issue No. 1
to set the first calibration value.
and F2
to set the calibration date.
DPI 740 User Manual