DPI 740 User Manual
Entering Local Height and Temperature Data for Sea
Level Pressure Measurements
The indicator stores the local height and temperature data
used in the sea level pressure measurement in nonvolatile
memory. To enter this data proceed as follows:
1.
Press the MODE push-button to select sea level pressure
measurement mode (confirmed by the display showing
"Sea" on the status line).
2.
Press F1 to show the measurement menu.
3.
Press F2 selecting the SET_QFF parameter (confirmed by
the display flashing "Enter Height ASL").
4.
Press either F1, F2 or SET to allow the height to be entered
(the display stops flashing).
5.
Use F1 to increase the height value and F2 to decrease
the height value.
Note:Pressing and holding either F1 or F2 changes the value
at a fast rate. Pressing and holding MODE and either F1 or F2
accelerates this fast rate.
6.
When the height is correct press SET, the display changes
to flash "Enter Temp".
7.
Repeat steps 4, 5 and 6 to enter the temperature value.
The display returns to sea level pressure measurement,
corrected using the new height and temperature data.
Note:Pressing the MODE push-button steps menu back one
setting, allowing a value to be corrected.