GE Druck DPI 740 User Manual page 60

Precision pressure indicator
Table of Contents

Advertisement

52
Command:
Format:
Description:
Query:
After entering all the points required to calibrate the
instrument, the accept calibration command CA must be sent
to update the data in the instrument's nonvolatile memory.
K0200 Issue No. 1
CP - Enter Calibration Point
CP = <value>
The calibration point value is given by
<value> in the currently selected units.
A query of this parameter CP? returns the
number of calibration points already
recorded by this command. The instrument
also indicates that it is ready for the next
calibration point.
Reply:
example
CP = 0
CP = 2
DPI 740 User Manual
no cal points recorded
2 cal points recorded