GE Druck DPI 740 User Manual page 62

Precision pressure indicator
Table of Contents

Advertisement

54
Command:
Format:
Description:
Example:
Query:
Example of a two point calibration
IU = 0
PP = 123
CT = 1
CP = 800
CP =1100
CA
CD
Command:
Format:
Description:
Example:
Query: There is no corresponding query command.
K0200 Issue No. 1
CD - Calibration date
CD = <dd/mm/yy>
Sets calibration date, only valid in calibration
mode.
CD = 24/01/97
set cal date to 24 January 1997.
CD?
query date of last calibration.
Reply:
CD = 24/01/97
set-up input as mbar
PIN entered put instrument
into calibration mode
perform two point calibration
applied pressure 800 mbar
applied pressure 1100 mbar
calibration accepted
calibration date
PP - PIN protection
PP = <pin>
This command is used to change the PIN of
the instrument which is used to protect the
calibration data and configuration.
<pin> = 123
PP= 123 PIN entered put instrument into
calibration mode
DPI 740 User Manual
calibration PIN code