Example:
Note:
This tares the value but does not change the displayed
value.
Example:
PC=<(IR)
Example:
PC=>(IR)
Example:
Example:
Define the derived ALTITUDE input value changing the
current or default setting of <datum pressure> to 120.00.
K0200 Issue No. 1
T
tare
PC=T(IR)
Full tare of input reading.
PC=T(IR,100.00)
Define tare of input reading, subtracts 100
from reading. The value specified in the
command is in the currently selected units.
<
minimum
Defines the minimum value of the input
reading. To reset this value use PM
command.
>
maximum
Defines the maximum value of the input
reading. To reset this value use PM
command.
Q
PC=Q(IR,200,20)
Define the derived QFF input value changing
both the <height> and <temp> parameters.
A
altitude (also see command PA)
PC=A(IR,120.00)