10.7.1

TCU requirements

TCU requirements - 50 L, 100 L,
200 L mixer
Property
General requirements
Process requirements
Remote control interface
(for XDUO mixers only)
TCU requirements - 500 L, 1000 L
mixer
Property
General requirements
Process requirements
Xcellerex XDM 50-1000, XDUO 100-1000 mixers Operating Instructions 29251615 AD
Parameter
Cooling capacity
Heating capacity
Fluid outlet pressure
Fluid outlet flow
Fluid outlet minimum temperature
Fluid outlet maximum temperature
Connector size
Setpoint
On/Off emergency stop
Parameter
Cooling capacity, 500 L mixer
Cooling capacity, 1000 L mixer
Heating capacity, 500 L and 1000 L mixer
Fluid outlet pressure
Fluid outlet flow
Fluid outlet minimum temperature
Fluid outlet maximum temperature
Connector size
10 Reference information
10.7 Temperature control unit
10.7.1 TCU requirements
Value
5 kW
5 kW
2.8 bar (40 psi g)
40 L/min
-10°C
65°C
1.5"
4 to 20 mA
Required
Value
9 kW
15 kW
9 kW
2.8 bar (40 psi g)
60 L/min
-10°C
65°C
1.5"
285