Prepare The Conductivity Probe - GE Xcellerex XDM Series Operating Instructions Manual

Table of Contents

Advertisement

6.4.2

Prepare the conductivity probe

Install the conductivity probe
To install and calibrate a new conductivity probe or to replace a probe, follow the instruc-
tions below.
Step
1
2
3
4
5
Xcellerex XDM 50-1000, XDUO 100-1000 mixers Operating Instructions 29251615 AD
Action
Take the probe out of the packaging.
Tip:
Keep the accompanying manual. You need it for calibration.
Rinse the sensor end of the probe with de-ionized water.
Connect the conductivity probe with the probe cable so that no thread is
visible on the probe.
NOTICE
Do not cross-thread the probe cable.
Find the cell constant value in the accompanying manual.
On the pH/conductivity transmitter display on the I/O Cabinet, press MENU
(1).
2
Result: A menu is displayed on the transmitter screen.
6.4 Prepare and install the probes
6.4.2 Prepare the conductivity probe
1
3
6 Preparation
163