Autoclave Probe Sheath Assembly - GE Xcellerex XDM Series Operating Instructions Manual

Table of Contents

Advertisement

6.4.3

Autoclave probe sheath assembly

Conditions
The probe sheath assembly with inserted probe must be autoclaved.
The recommended conditions for autoclaving are as follows:
Prepare for autoclaving
Before autoclaving the probe sheath assembly, prepare the following:
Follow the instructions below to prepare the probe sheath for autoclaving:
Step
1
2
Xcellerex XDM 50-1000, XDUO 100-1000 mixers Operating Instructions 29251615 AD
Temperature > 121°C
Time 60 min (minimum time 30 min)
Liquid cycle
NOTICE
The autoclaving temperature must not exceed 130°C.
NOTICE
Do not autoclave ratchet clamps.
Disposable pipette
Deionized water
Action
Verify that the probe cable is not connected to the probe.
Draw 1 to 2 mL deionized water into a disposable pipette.
6.4 Prepare and install the probes
6.4.3 Autoclave probe sheath assembly
6 Preparation
165