6.4.3
Autoclave probe sheath assembly
Conditions
The probe sheath assembly with inserted probe must be autoclaved.
The recommended conditions for autoclaving are as follows:
•
•
•
Prepare for autoclaving
Before autoclaving the probe sheath assembly, prepare the following:
•
•
Follow the instructions below to prepare the probe sheath for autoclaving:
Step
1
2
Xcellerex XDM 50-1000, XDUO 100-1000 mixers Operating Instructions 29251615 AD
Temperature > 121°C
Time 60 min (minimum time 30 min)
Liquid cycle
NOTICE
The autoclaving temperature must not exceed 130°C.
NOTICE
Do not autoclave ratchet clamps.
Disposable pipette
Deionized water
Action
Verify that the probe cable is not connected to the probe.
Draw 1 to 2 mL deionized water into a disposable pipette.
6.4 Prepare and install the probes
6.4.3 Autoclave probe sheath assembly
6 Preparation
165