Functional Test Limits For Accelerometers And Velocimeters; Using Model 11A Calibrator; Balance Measure Only Menu - Honeywell VXP System Maintenance Manual

Vibration analysis system
Table of Contents

Advertisement

7.
In the Bal Meas Only menu (Figure 4-3), enter the following settings:
Select the AZIMUTH 1 in the AZ CH: pull down window.
Select the VELO 1 in the VIB CH: pull down window.
Enter 900 in the BAL RPM: window by highlighting the existing entry on the
display with the stylus, and then typing in the number using the keypad.
Select 1 in the N PER REV pull down window.
Select AUTO STOP in the pull down window.
8.
Click on the START button.
9.
Verify that the correct RPM is displayed in the upper left corner of the Bal Meas
Only window.
10. Verify that the correct magnitude and phase readings are displayed, depending upon
the type of sensor used. Table 4-8 lists the correct readings for Chadwick-Helmuth
accelerometers and velocimeters.
11. Repeat the preceding steps until all eight of the discrete Velocimeter/Accelerometer
inputs and all of the sensors and cables have been tested.
Table 4-8. Functional Test Limits for Accelerometers and Velocimeters
Sensor Type/
Part No.
Accelerometer (ICP)
906-13728
906-13925
906-13955
Accelerometer (High
Temp)
906-7569A
Figure 4-3. Balance Measure Only Menu

using Model 11A Calibrator

Sensitivity
RPM
10 mV/G
900
100 pC/G
900
Magnitude
Phase
(ips)
(hrs:min)
1.00 ±0.15
6:00 ±0:30
0.055 ±0.008
6:00 ±0:30
VXP System Maintenance Manual
Maintenance
4-7