Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors
3.3 5-Module Frame disassembly and reassembly
3.3.1 Before disassembly
WARNING
Perform a leakage current measurement whenever service or repair has
been done on the device.
WARNING
Wear a static control wrist strap when handling PC boards. Electrostatic
discharge may damage components on the board.
Disconnect the main power cord. If the monitor is connected to the Datex-Ohmeda Network,
disconnect the Mon-Net cable.
3.3.2 Tools needed
3.3.3 To disassemble the 5-Module Frame, F-CU5(P)
26
.
Document no
M1125636-05
-
screwdriver, TORX; T10, T20
-
flat blade screwdriver
-
antistatic wristband
-
monkey wrench, size 5.5
1. Remove all the cables from the F-CU5(P).
2. Remove the screw (T10) to open the back cover
(including fan filter). Press the locking clamps to
remove the back cover.