Network Config - GE Datex-Ohmeda S/5 Technical Reference Manual

Anesthesia monitor, critical care monitor
Table of Contents

Advertisement

Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors
Subnet Status
The Subnet status view gives more accurate
information of the different subnet id:s connected. All
four Subnet status menus have a similar structure.
The number of different packets transmitted and
received by the monitor are shown in the columns
below Tx and Rx. The packet types are:
Waveforms: Waveform data
Phys. data:
Physiological numerical data
Alarms:
Alarms, alarm profiles and alarm limits
Link mgmt:
Network management messages
Record K:
Record Keeper data
MonToMon:
Monitor-to-monitor communication
related data
Printer:
Printing data and control messages
File Op.:
File operation messages, saving and
loading of cases
Service:
Maintenance and service
Modes:
User mode data
Indics.:
Remote indications sent to monitor
RemoteEv:
Remote events
Data server: Packets of the data server (Arrhythmia Workstation)
Packets total: Total number of packets sent/received
Bytes total:
Total number of bytes sent/received
T-o
Number or resendings
InE
Received faulty packets
LenE
Erroneous packet length
Dupl
Same packet received as a duplicate

1.3.2 Network Config

The DRI Level is for setting the monitor's network
communication. The network communication is
set according to the network software used (e.g.
S-CNET01).
Network software S-CNET01 ->DRI Level = 2001
Network software S-CNET02 ->DRI Level = 2001
Network software L-NET03 ->DRI Level = 2003
Network software L-NET05 ->DRI Level = 2005
Network software L-NET05-02 ->DRI Level = 2008
6
.
Document no
M1137266-02

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents