Equipment Safety Symbols - GE Datex-Ohmeda S/5 Technical Reference Manual

Anesthesia monitor, critical care monitor
Table of Contents

Advertisement

Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors

1.1.3 Equipment safety symbols

6
.
Document no
M1137263-04
-
Attention, consult accompanying documents.
-
When displayed next to the O
is set below 21%.
-
When displayed next to the HR value, indicates that the pacer is set on R.
-
On the modules or frames indicates the following warning:
WARNING
Do not use modules with identical measurements in
the same monitor. If such modules have been inserted,
remove the module that has been most recently
connected. You can also remove both modules and
re-connect the new module after five seconds.
-
On the 15" Flat Panel Display, D-FPD15-00, indicates the following warning:
WARNING
The display must only be used together with the
original type of D-FPD15 power adapter. The display
should be supplied from the mains outlet.
-
On the 19" display, D-LCC19, indicates the following warning:
WARNING
The display must only be supplied from the mains
outlet via an appropriate additional separating
transformer and the original D-LCC19 power adapter,
not from the Central Unit, F-CU8.
-
On the E-PRESTN, E-PRETN, E-RESTN, E-PSM, E-PSMP, E-P, E-PP, E-PT, E-COP
and E-COPSv module indicates the following warning:
WARNING
Protection against cardiac defibrillator discharge is
due in part to the accessories for pulse oximetry
(SpO2), temperature (T) and invasive pressure (P)
measurement.
-
On the E-NMT module indicates the following warnings:
WARNING
Do not place the NMT stimulating electrodes on the
patient's chest.
WARNING
Always stop the NMT measurement before handling
the stimulating electrodes.
WARNING
Never subject a patient with an implanted electronic
device to electrical stimulation without consulting a
medical specialist first.
-
On the rear or bottom panel this symbol indicates the following warnings
and caution:
WARNING
Electric shock hazard. Do not open the cover or the
back. Refer servicing to qualified service personnel.
WARNING
Disconnect from the power supply before servicing.
WARNING
Do not use the monitor without manufacturer
approved mounting attached.
CAUTION
For continued protection against fire hazard, replace the
fuse only with one of the same type and rating.
value, indicates that the FiO
2
low alarm limit
2