Datex-Ohmeda S/5 Anesthesia and Critical Care Monitors
Perform a NIBP measurement and check that the module gives a reasonable measured
result.

5.3.9 Pressure/Pressure Temp Modules, E-P, E-PT

Invasive blood pressure measurement
1.
Check the function of the front panel
2.
Check the InvBP channel with a patient simulator.
Zero the InvBP channel. Then check that the values and waveforms correspond to the
simulator settings.
Temperature measurement
3.
Check the temperature channels with a patient simulator.
Check that the temperature measurement information is shown on the monitor screen as
configured on the simulator.

5.3.10 Dual pressure Module, E-PP

1.
Check the function of the front panel
2.
Check the InvBP channels with a patient simulator.
Zero the InvBP channels. Then check that the values and waveforms correspond to the
simulator settings.

5.3.11 Cardiac Output Modules, E-COP, E-COPSv

Invasive blood pressure measurement
1.
Check the function of the front panel
2.
Check the InvBP channel with a patient simulator.
Zero the InvBP channel. Then check that the values and waveforms correspond to the
simulator settings.
Cardiac Output measurement
3.
Check the function of the front panel
70
.
Document no
M1137263-04
Zero P3
key.
Zero P5
Zero P6
and
keys.
Zero P4
key.
Start C.O.
key.