4.1.2 Connection to Critikon Dinamap 1846SX, Abbott Oximetrix 3 and Baxter
Explorer
Use the INT-External Device Cable.
1.
Make sure that the power to both monitors is turned off.
2.
Connect the 9 pin D-connector to the Interface Module, E-INT. Tighten the finger screws.
3.
Connect the 25 pin D-connector to the connector on the external monitor. Tighten the
finger screws.

4.1.3 Connection to Baxter Vigilance

Use the INT-Baxter Vigilance Cable.
1.
Make sure that the power to both monitors is turned off.
2.
Connect the 9 pin D-connector to the Interface Module, E-INT. Tighten the finger screws.
3.
Connect the other 9 pin D-connector to a corresponding connector on the external
monitor. Tighten the finger screws.
Communication parameters are: baud rate 19200, no parity (none), data bits 8, stop bits 1.
Mode of communication port: IFM out.

4.1.4 Connection to Nellcor N-100 and N-1000

Use the Monitor-Nellcor Cable.
1.
Make sure that the power to both monitors is turned off.
2.
Connect the 9 pin D-connector to the Interface Module, E-INT. Tighten the finger screws.
NOTE: The X3 and X9 connectors cannot be used when interfacing the S/5 Anesthesia Monitor
or the S/5 Critical Care Monitor with Nellcor N-100 monitors.
3.
Connect the other connector to the corresponding connector on the external monitor.
Tighten the finger screws. Connect also the mono connectors on the external monitor.

4.1.5 Connection to Nellcor N-200

Use the Monitor-Nellcor Cable.
1.
Make sure that the power to both monitors is turned off.
2.
Connect one of the connectors to the corresponding connector on the Interface Module,
E-INT. Tighten the finger screws.
3.
Connect the other connector to the corresponding connector on the external monitor.
Tighten the finger screws. Connect also the mono connectors on the external monitor.
Connect also the mono connectors on the external monitor.
Interfacing
Document no. M1137263-04
49