Overview - ABB ProcessMaster FEP611 Operating Instructions Manual

Table of Contents

Advertisement

8.3

Overview

Errors encountered are itemized in tabular form on the following pages. The response of the transmitter on error detection is
described therein. The table lists all possible errors together with a description of their impact on the value of measurement
variables, the properties of current outputs and the alarm output.
Error no. /
Error text
Range
F099.042 /
ADC overrange alarm.
Electronics
Noise too high. Check
applicati. Call service.
F098.012 /
FEB communication error.
Electronics
EMC disturbance. Call ABB
Service.
F097.029 /
Coil regulation error.
Electronics
Check wiring of sensor coils.
Call service
F096.043 /
SIL self check alarm.
Electronics
Call service.
F095.036 /
Coil isolation alarm.
Electronics
Call service.
F094.022 /
Safety Alarm Curr. Out 31 /
Electronics
32
SIL function detects error.
Call Service.
F093.032 /
Electrode short cuircit.
Check wiring of sensor
Electronics
electrode. Call service.
F092.033 /
Electrode open cuircit.
Check wiring of sensor
Electronics
electrode. Call service.
F091.030 /
Coil wiring error.
Electronics
Check wiring of sensor coils.
Call service.
F090.035 /
ADC RX210 com. error.
Call service.
Electronics
F088.013 /
FEB communication error.
Electronics
EMC disturbance. Call ABB
Service.
F086.019 /
Curr.Out 31 / 32 com error.
Electronics
Defective Board. EMC
disturbance. Call Service.
F084.011 /
NV data defect.
Electronics
Data storage irreparable.
Call Service.
F082.014 /
Incompatible Frontend
Electronics
Board.
Frontend not fit to
Motherboard. Call Service.
F081.026 /
MB voltages outside range.
Electronics
Defective Motherboard HW.
Call Service.
68 OI/FEP610/FEH610-EN Rev. B | ProcessMaster FEP610, HygienicMaster FEH610
Current output
Digital output
High Alarm or Low
General Alarm, if
Alarm, depending
DO is configured
on parameter
as "Logic / Alarm
"Curr.Out at
Signal"
Alarm".
Pulse output
LCD display
0 Hz
0 %
0 Hz
0 %
0 Hz
0 %
Current value -
Current value -
no change.
no change.
Current value -
Current value -
no change.
no change.
Current value -
Current value -
no change.
no change.
0 Hz
0 %
0 Hz
0 %
0 Hz
0 %
0 Hz
0 %
0 Hz
0 %
Current value -
Current value -
no change.
no change.
0 Hz
0 %
0 Hz
0 %
0 Hz
0 %
Error maskable?
No
No
No
No
No
No
No
No
No
No
No
No
No
No
No

Hide quick links: