Error Messages; Errors - ABB ProcessMaster FEP611 Operating Instructions Manual

Table of Contents

Advertisement

8.2

Error messages

The error messages are divided into four groups in accordance with the NAMUR classification scheme.
8.2.1

Errors

Error no. / Range
Text on the LCD display
F099.042 / Electronics
ADC overrange alarm.
Noise too high. Check
applicati. Call service.
F098.012 / Electronics
FEB communication error.
EMC disturbance. Call ABB
Service.
Coil regulation error.
F097.029 / Electronics
Check wiring of sensor coils.
Call service
F096.043 / Electronics
SIL self check alarm.
Call service.
F095.036 / Electronics
Coil isolation alarm.
Call service.
Safety Alarm Curr. Out 31 /
F094.022 / Electronics
32
SIL function detects error.
Call Service.
F093.032 / Electronics
Electrode short cuircit.
Check wiring of sensor
electrode. Call service.
F092.033 / Electronics
Electrode open cuircit.
Check wiring of sensor
electrode. Call service.
F091.030 / Electronics
Coil wiring error.
Check wiring of sensor coils.
Call service.
F090.035 / Electronics
ADC RX210 com. error.
Call service.
F088.013 / Electronics
FEB communication error.
EMC disturbance. Call ABB
Service.
Cause
Electrode signal exceed max ADC Limits.
No flow measurement possible.
Defective Frontend Board or Motherboard
Hardware.
Incorrect wiring between sensor and
transmitter. Wrong or no connection between
those two boards.
Incorrect coil wiring (terminals M1 / M2) or
Wire break / short circuit or defect of coil
drive fuse or moisture in sensor terminal box.
The SIL monitoring function has detected a
transmitter error.
Defective coil incorrect wiring (shortcut
M1 / M2 vs. GND). Flooded sensor
The Current Output μController detects SIL
relevant errors.
Incorrect wiring or electrode leakage or short
cut of electrode signal lead vs. shield or
sensor flooded.
Incorrect electrode wiring or electrode signal
lead break.
Incorrect coil wiring (terminals M1 / M2) or
Wire break / short circuit or defect of coil
drive fuse or moisture in sensor terminal box.
Bad EMC environment or defective
component.
EMC disturbance on the signal cable.
Wrong signal cable.
ProcessMaster FEP610, HygienicMaster FEH610 | OI/FEP610/FEH610-EN Rev. B 65
Remedy
In case of an empty pipe make sure the
empty pipe detector is enabled.
Make sure actual flowrate does not exceed
flow range configured.
Call Service
With integral design: Call service.
With remote design: Check for wiring fault
between sensor and remote transmitter.
Call service
Check for incorrect coil wiring (terminals
M1 / M2) or Wire break / short circuit or
defect of coil drive fuse or moisture in sensor
terminal box.
Call service
Call service
Call service
Check for incorrect wiring or electrode
leakage or short cut of electrode signal lead
vs shield or sensor flooded.
Check for incorrect electrode wiring or
electrode signal lead break.
Check for incorrect coil wiring (terminals
M1 / M2) or Wire break / short circuit or
defect of coil drive fuse or moisture in sensor
terminal box.
Replace electronic or Contact Service.
Check Signal cable and connection.
Call Service.