Table 6-8 Input 1 Calibration Procedure - Honeywell UDC2300 Product Manual

Universal digital controller
Hide thumbs Also See for UDC2300:
Table of Contents

Advertisement

Input Calibration
Procedure
The calibration procedure for Input #1 is listed in Table 6-8.
Step
Operation
1
Enter Calibration
Mode
Calibrate 0 %
2
Calibrate 100 %
3
50

Table 6-8 Input 1 Calibration Procedure

Press
Upper Display = CAL
SET UP
Lower Display = INPUT1
until you see
You will see:
FUNCTION
Upper Display = DIS
Lower Display = CALIN1
The calibration sequence is enabled and you will see:
Upper Display = BEGN
Lower Display = CALIN1
At the completion of the sequence, the selection
automatically reverts to disable.
You will see:
FUNCTION
Upper Display = APLY
Lower Display = IN1ZRO
Adjust your calibration device to an output signal equal to
the 0 % range value for your particular input sensor. See
Table 6-1 for Voltage, Degrees, or Resistance
equivalents for 0 % range values.
Wait 15 seconds, and then go to the next step.
You will see:
FUNCTION
Upper Display = APLY
Lower Display = IN1SPN
Adjust your calibration device to an output signal equal to
the 100 % range value for your particular input sensor.
See Table 6-1 for Voltage, Degrees, or Resistance
equivalents for 100 % range values.
Wait 15 seconds, and
If ...
you are calibrating a Thermocouple input go to step 4
you are calibrating other than a
Thermocouple input
UDC2300 Limit Controller Product Manual
Result
Then ...
go to step 5
1/01