Measurements and Patient Worksheets (Reports)
Note
Note
11-4
To delete all measurement results of the selected "Study" from the monitor as well as from the
corresponding Worksheet, select the [Clear Study] item on the menu area.
When [Clear Study] runs, it shows below message.
To erase measurement results:
Select the [Clear Study] item on the menu area
or select the [Delete Last] on the menu area
To exit the Generic Measurement program:
Press the [Caliper] key on the control panel
To get optimum resolution and accuracy from Doppler measurements, the [Angle]
correction cursor must be positioned parallel to the vessel axis (in the area of the
measuring volume).
When the result display is full, (max. 4) the first measurement will be overwritten first.
Except for Auto Trace measurements, all measurement results will be automatically
included in the corresponding Generic Worksheet. To store Auto Trace measurement
results, press the right or left trackball key [Set] previously.
Depending on the Application setting and the adjustment in the Measure Setup:
RI and PI will be calculated using ED (End Diastole) or "MD" (Mid Diastole)
Vdiastole = Vend-diastole or Vmin (depending on this selection)
all previously set measuring marks are erased when starting a new scan (unfreeze
-> Run mode), or the measurement results are kept on screen.
the Spectral Doppler envelope curve is performed with a continuous trace line or
by setting points
the Doppler measuring results (according to the "Auto/Manual Trace" setting) are
displayed after an Auto- or Manual Trace measurement
For further details, review: Application Parameters
page 11-109 )
(chapter 'The Measure Setup Pages' on