3 S y s t e m C o n t r o l s

Image Task Card

The Image task card is used for selecting the imaging mode and the
image format and for optimizing the settings such as dynamic range,
persistence, and edge enhancement or for selecting an exam type or
transducer that contains pre-configured, optimized settings.
The Image task card also provides controls for functions other than imaging
parameters. VCR recording, ECG settings, image formatting, and the update
styles settings are organized in group boxes.
The Image task card also includes capture definitions for storing and
reviewing clips.

Parameter Menu Selections

Mode-specific menus and selections display on the left side of the screen in
the Image task card. Each imaging mode has a mode-dependent Parameter
menu used to adjust imaging parameters.
Note: A Parameter menu may have two pages of selections. To access the menu
selections, roll the trackball to the Page 1 of 2 indicator or the Page 2 of 2 indicator on the
Parameter menu and then press the SELECT key.
Note: A menu selection can be an icon and/or text. The "Tool Tip" column in the following
tables indicates additional text displayed on the screen when you roll the trackball to
position the trackball pointer on the menu selection. The "Settings" column in the
following tables lists the settings that are displayed on the Parameter menu for each
menu selection (graphically depicted settings or options are not listed in this column).

2D-Mode Parameter Menu Selections

Menu
Selection
Tool Tip
(current
Transmit
setting)
Frequency
#Foc
# Focal Points
DR
Dynamic Range
R/S
Res/Speed
Edge
Edge Enhance
3 - 2 6
Description
Changes the transmit frequency of an active
multi-frequency transducer in 2D-mode.
Selects the number of focal zones.
Controls the overall contrast resolution of the image.
Adjusts the balance between the image line density
(resolution) and the frame rate.
Distinguishes the contours of a structure during
real-time imaging.
[1] Instructions for Use
Protocol Group Box 3-38
Settings
Transducer-
dependent
Dependent on
transducer and
exam type
30 dB to 70 dB in
5 dB increments
0 to 5
0 (none), 1 to 3
[ 1 ]
I N S T R U C T I O N S