Siemens ACUSON Antares Instructions For Use Manual page 113

Ultrasound imaging system
Table of Contents

Advertisement

2D-Mode Parameters for
Note: A Parameter menu may have two pages of selections. To access the menu
selections, roll the trackball to the Page 1 of 2 indicator or the Page 2 of 2 indicator on the
Parameter menu and then press the SELECT key.
Menu
Selection
Tool Tip
(current
Rendering Mode
setting)
Plane
Cut Plane
Maps
Maps
DR
Dynamic Range
Tint
Tint
Low Th
Low Threshold
High Th
High Threshold
[ 1 ]
I N S T R U C T I O N S
F O R
four
Sight Imaging
Description
Renders the volume using the selected rendering
method:
Opacity—Smoothes image contours, creating a soft,
sculptured appearance for highlighting surface
features. Uses the selections for opacity percentage,
thresholds, and smoothing.
MinIP (Minimum Intensity Projection)—Limits the
display to minimum-value pixels by selecting pixels
with the lowest value along each projection of the
volume. May be helpful in visualizing hyper-echoic
structures and fetal skeletons.
MaxIP (Maximum Intensity Projection)—Limits the
display to maximum-value pixels by selecting pixels
with the highest value along each projection of the
volume. May be helpful in visualizing hypo-echoic
lesions and large vascular structures.
MeanIP (Mean Intensity Projection)—Limits the
display to average-value pixels by selecting pixels
with the arithmetic mean (average) value along each
projection of the volume.
Enables or disables the specified cut plane for 2D-mode
data in the volume.
Note: This selection is not available when VOI is enabled.
Selects a processing curve that assigns echo
amplitudes to gray levels.
Note: Applies to slice quadrants only.
Controls the overall contrast resolution.
Applies the selected tint to the 2D-mode data.
Sets the low threshold for the opacity curve. Removes
2D-mode data (voxels) that have values lower than the
selected threshold.
Sets the high threshold for the opacity curve. Assigns a
constant opacity value to all voxels at or above the
selected threshold.
Note: Use Opacity instead of High Th to adjust image
transparency.
U S E
3 S y s t e m C o n t r o l s
Settings
Opacity
MinIP
MaxIP
MeanIP
Off, 1, 2
A, B, C, D, E,
F, G, H, I
Transducer-
dependent
Transducer-
dependent
0 to 254
1 to 255
3 - 4 1