2 S a f e t y a n d C a r e
3. Remove the air filter from the tray.
4
3
2
1
Removing/replacing the air filter.
4. Rinse the air filter with running water and allow the filter to
completely dry.
To hasten drying, you may gently shake the filter, or blot the filter with a
clean, lint-free cloth.
Caution: Do not insert a wet filter as this can damage the system.
5. Replace the air filter into the tray, positioning the filter flat side down,
with the angled corners fitting into the back of the tray. The filter will fit
into the back corners of the tray only if the filter is positioned flat
side down.
6. Press the filter against the hook and loop fasteners on the tray.
7. Slide the air filter tray back into the ultrasound system.
8. Plug the power cord into the power outlet.
2 - 3 6
1 Air filter tray
2 Hook and loop
fasteners
3 Air filter
4 Flat side of air
filter
[ 1 ]
I N S T R U C T I O N S
F O R
U S E