Power Mode-Specific Selections For 3-Scape Imaging - Siemens ACUSON Antares Instructions For Use Manual

Ultrasound imaging system
Table of Contents

Advertisement

3 S y s t e m C o n t r o l s
Power Mode-Specific Selections for
3-Scape Imaging
Menu
Selection
Tool Tip
(current
Color Rendering
setting)
Mode
Plane
Color Cut Plane
Maps
Maps
Priority
Tissue Reject
(Priority)
Low Th
Low Threshold
3 - 5 6
Description
Renders the Power mode data in the volume using the
selected rendering method:
Opacity—Smoothes image contours, creating a soft,
sculptured appearance for highlighting surface
features. Uses the selections for opacity percentage.
MaxIP (Maximum Intensity Projection)—Limits the
display to maximum-value pixels by selecting pixels
with the highest value along each projection of the
volume. May be helpful in visualizing hypo-echoic
lesions and large vascular structures.
Surf. Shading—Heightens contrast, creating a
textured appearance for highlighting internal features.
Uses the selected shading, opacity percentage, and
thresholds.
Note: Surf. Shading is not available for volumes
acquired using the Auto Sweep method.
Enables or disables the specified cut plane (1 or 2) for
Power data in the volume.
Note: This selection is not available when VOI is enabled.
Selects a processing curve that assigns flow amplitudes
to color levels.
Adjusts the tissue rejection level threshold for the
amplitude of the Power display.
Sets the low threshold for the opacity curve. Removes
Power-mode data (voxels) that have values lower than
the selected threshold.
Settings
Opacity
MaxIP
Surf.
Shading
Off, 1, 2
For slices:
A, B, C, D, E,
F, G, H
For the
volume:
A, B, C, D, E,
F, G, H, I, J, K
0 to 15
0 to 254
[ 1 ]