Transducer Surface Temperature Limits

The following table provides the maximum surface temperature of the
transducers compatible with the system.
Maximum surface temperatures are in accordance with IEC 60601-2-37.
Transducer
C5F1
C7F2
C5-2
CX5-2
CH4-1
CH6-2
VF7-3
VFX9-4
VF10-5
VF13-5
VF13-5SP
VFX13-5
P10-4
PH4-1
PX4-1
EC9-4
EV9F4
TMM = Tissue Mimicking Material
[ 1 ]
I N S T R U C T I O N S
F O R
Maximum Temperature
TMM
Still Air
≤41.0°C
≤47.0°C
≤41.0°C
≤45.5°C
≤41.0°C
≤41.5°C
≤41.0°C
≤43.5°C
≤41.5°C
≤47.0°C
≤41.0°C
≤41.0°C
≤41.0°C
≤46.0°C
≤41.5°C
≤47.5°C
≤41.0°C
≤46.0°C
≤41.0°C
≤41.5°C
≤41.0°C
≤41.5°C
≤41.0°C
≤42.0°C
≤41.0°C
≤41.0°C
≤41.0°C
≤41.0°C
≤41.0°C
≤43.0°C
≤41.5°C
≤41.0°C
≤41.0°C
≤41.0°C
U S E
2 S a f e t y a n d C a r e