Potentiometers and Reference Switches
Basic information on the replacement and adjustment of
potentiometers, reference switches and motor encoders
Fig. 4:

General replacement and adjustment

a) Potentiometers
Move the unit so that the set screw becomes accessible
Switch the system off.
Remove the Allen 1.5 mm set screw on the connecting piece of the potentiometer shaft
((M/Fig. 4 / p.
Remove the potentiometer retaining plate
Remove the adjustment wheel
tentiometer shaft.
Desolder the connection wires from the potentiometer and solder them to the new po-
tentiometer in the same order.
Install the potentiometer in the reverse order. Do not tighten the set screw yet (Allen 1.5
mm).
Switch the system on.
Move the drive in "unregulated motion" with the service PC (Adjustment, Calibration) in
mm steps so that the corresponding reference switch is actuated (refer to
(Fig. 5 / p.
NOTE
Exit "unregulated motion" and select the "ADC value" for the corresponding drive in Di-
agnostic, Monitoring.
Siemens AG
Medical Solutions
Replacing the potentiometer
35)).
36)). Check with DVM at D5.X2.A1...A11, 0V=X30.
The potentiometer values are set in accordance with ADC values.
RXL2-120.841.02.01.02
09.05
((ISK 3 mm/Fig. 4 / p.
((E/Fig. 4 / p.
35)) and the retaining parts from the po-
Page 35 of 72
CS PS 24
((M/Fig. 4 / p.
35)).
35)).
LITHOSTAR Multiline
35