SIREMOBIL Iso-C
Service Instructions
Replacing the horizontal lift brake
Print No.: SPR2-230.076.01.01.02
Replaces: n.a.
SP
© Siem ens A G 1 999
The reproduction, transmission or
use of this document or its contents
is not permitted without express
written authority. Offenders will be
liable for damages. All rights,
including rights created by patent
grant or registration of a utility
model
_or_ design,_are_
reserved.
English
Doc. Gen. Date: 02/99

Advertisement

loading

Summary of Contents for Siemens SIREMOBIL Iso-C

  • Page 1 SIREMOBIL Iso-C Service Instructions Replacing the horizontal lift brake © Siem ens A G 1 999 The reproduction, transmission or use of this document or its contents is not permitted without express written authority. Offenders will be liable for damages. All rights,...
  • Page 2 These instructions describe the replacement of the horizontal lift brake on the SIREMO- BIL Iso-C. Required parts • Magnetic brake (Item number: refer to the SIREMOBIL Iso-C spare parts list) Required measurement and auxiliary devices • Protective conductor meter 44 15 899 RV090 •...
  • Page 3: Preliminary Steps

    C-arm is positioned flat on the table with the entire rail- ing. Fig. 1 • Connect the SIREMOBIL Iso-C to power . • Position the wooden block (1/Fig. 1) underneath. • Move the C-arm into the horizontal position (Fig. 1).
  • Page 4 Remove the guide bearings (1/Fig. 5). • Remove plug connections X22 and X23 (2/Fig.5). • Remove the cable clamps (3/Fig.5). • Remove the cable guide (4/Fig.5). SIREMOBIL Iso-C SPR2-230.076.01 Page 4 of 16 Siemens AG Rev. 01 02/99 TD SD 24...
  • Page 5 When doing this, make sure to hold onto the 3 screws of the angulation brake or use adhesive tape to prevent them from falling. Fig. 9 shows the complete angulation brake. Siemens AG SPR2-230.076.01 Page 5 of 16 SIREMOBIL Iso-C Medical Engineering Rev. 01 02/99 TD SD 24...
  • Page 6 • Once the end stop has been removed, the shaft should lie horizontal within the horizontal carriage and should not be pressing against it. • Switch OFF the SIREMOBIL. SIREMOBIL Iso-C SPR2-230.076.01 Page 6 of 16 Siemens AG Rev. 01...
  • Page 7 Remove the spring ring (1/Fig.12). • Remove the stop wheel (1/Fig.12). • Remove the end stop (1/Fig.13). • Remove the spring ring (2/Fig.13). Siemens AG SPR2-230.076.01 Page 7 of 16 SIREMOBIL Iso-C Medical Engineering Rev. 01 02/99 TD SD 24...
  • Page 8 Replacing the horizontal lift brake Fig. 14 Fig. 15 • Switch ON the SIREMOBIL Iso-C and release the horizontal brake (the LED in the switch lights). • Move the horizontal carriage so that the opening in the horizontal carriage, (1/Fig.14) as shown in Fig.14 is located opposite the opening in the bearing support (2/ Fig.14).
  • Page 9: Installation

    (1/Fig.18, 1/Fig.19). • Route the cable end of the horizontal brake parallel to the cable bundle in the horizontal carriage. Siemens AG SPR2-230.076.01 Page 9 of 16 SIREMOBIL Iso-C Medical Engineering Rev. 01 02/99 TD SD 24...
  • Page 10 - When inserting the horizontal carriage, keep it centered to the ball bearings. - You may have to adjust the position of the vertical lifting column until the horizontal carriage moves easily. SIREMOBIL Iso-C SPR2-230.076.01 Page 10 of 16 Siemens AG Rev.
  • Page 11 Reinstall the spring ring (1/Fig.21). • Reattach the stop (2/Fig.21) with the screw. • Reinstall the stop wheel (1/Fig.22). • Reinstall the spring ring (2/Fig.22). Siemens AG SPR2-230.076.01 Page 11 of 16 SIREMOBIL Iso-C Medical Engineering Rev. 01 02/99 TD SD 24...
  • Page 12 (Fig. 26 indicates the schematic top view of the horizontal carriage).You may have to adjust the eccentric bolt (2/Fig. 25). SIREMOBIL Iso-C SPR2-230.076.01 Page 12 of 16 Siemens AG Rev.
  • Page 13 When doing this, move the horizontal carriage several times from end stop to end stop to eliminate the play on the shaft. Siemens AG SPR2-230.076.01 Page 13 of 16 SIREMOBIL Iso-C Medical Engineering Rev. 01 02/99 TD SD 24...
  • Page 14 (Refer to Fig. 30). • Do not attach the cable for the brake switches (connector X1) with the other cables in the cable clamp (1/Fig. 31). SIREMOBIL Iso-C SPR2-230.076.01 Page 14 of 16 Siemens AG Rev.
  • Page 15 Route the cables to microswitches S3 and S4 and connect them (2/Fig. 31). • Secure the cables to the microswitches to the side of the plate with cable ties (2/Fig. 31). Siemens AG SPR2-230.076.01 Page 15 of 16 SIREMOBIL Iso-C Medical Engineering Rev. 01 02/99 TD SD 24...
  • Page 16: Concluding Work

    ARTD-002.731.17. • Perform a function test on the SIREMOBIL. TD SD 24 / Arnold TD SD 24 / Groß SMS Iselin / O’Donnell SIREMOBIL Iso-C SPR2-230.076.01 Page 16 of 16 Siemens AG Rev. 01 02/99...