Overview of Measures

VIDEOMED DI-M component

When servicing or replacing the VIDEOMED DI-M, adhere to the Service Instructions
RX52-022.061.01...
Board
Part no.
D1
30 97 727
X2170
D3
30 97 651
X2170
Siemens AG
Medical Solutions
Name
Remarks regarding replacement / tool tips
Analog board Remove the resistor between X902 and X903,
refer to the service note "Adapting a
MEMOSKOP"
Digital
"Load Default Value" is possible the first time only
with the switch, not with the service software.
control
Select the LITHOSTAR measurement field;
board
perform a dose rate check;
perform an IQ test
For 33 cm I.I. only:
select horizontal and vertical image reversal
RXL2-120.841.02.01.02
09.05
Page 13 of 72
CS PS 24
13
0