Siemens LITHOSTAR Multiline Maintenance Instructions Manual
Siemens LITHOSTAR Multiline Maintenance Instructions Manual

Siemens LITHOSTAR Multiline Maintenance Instructions Manual

Hide thumbs Also See for LITHOSTAR Multiline:

Advertisement

Quick Links

E r l w e i n
LITHOSTAR Multiline
Maintenance Instructions
System
LITHOSTAR Multiline
The protocol RXL2-120.105.01.06.02 is required for
these instructions
Print No.:
RXL2-120.101.01.06.02
Replaces: RXL2-120.101.01.05.02
03127011
SP
1995
© Siemens AG
The reproduction, transmission or use
of this document or its contents is not
permitted
without
express
authority. Offenders will be liable for
damages. All rights, including rights
created by patent grant or registration
of a utility model or design, are
reserved.
English
Doc. Gen. Date: 08.05
written

Advertisement

Table of Contents
loading

Summary of Contents for Siemens LITHOSTAR Multiline

  • Page 1 E r l w e i n LITHOSTAR Multiline Maintenance Instructions System LITHOSTAR Multiline The protocol RXL2-120.105.01.06.02 is required for these instructions 03127011 1995 © Siemens AG The reproduction, transmission or use of this document or its contents is not...
  • Page 2 Assemblers and other persons who are not employed by or otherwise directly affiliated with or authorized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or one of its affiliates before attempting installation or service pro- cedures.
  • Page 3: Table Of Contents

    Generator ............34 Siemens AG RXL2-120.101.01.06.02 LITHOSTAR Multiline Page 3 of 52 Medical Solutions 08.05...
  • Page 4 Unit function ............50 13 ______ Changes to Previous Version _____________________________________ 51 LITHOSTAR Multiline RXL2-120.101.01.06.02...
  • Page 5: Required Documents

    • Planning RXL2-120.021.. • Maintenance Protocol RXL2-120.105.01.05. • Image quality binder RXD0-000.064.01.. • Ultrasound Localizer Start-up RXL2-120.091.10... • Shock Wave Pressure and Position Check SPL2-120.074.01... Siemens AG RXL2-120.101.01.06.02 LITHOSTAR Multiline Page 5 of 52 Medical Solutions 08.05 CS PS 24...
  • Page 6: Test And Auxiliary Equipment

    34 64 062 • TV dynamic test case 37 90 156 X 1963 • PTU pressure tester 30 95 408 - PTU adapter 70 41 028 LITHOSTAR Multiline RXL2-120.101.01.06.02 Siemens AG Page 6 of 52 08.05 CS PS 24 Medical Solutions...
  • Page 7: Tools

    44 30 906 RH 090 • for POLYDOROS: Installation material "washers" 88 67 665 X 1122 • 1 liter transformer oil as needed 99 00 945 Siemens AG RXL2-120.101.01.06.02 LITHOSTAR Multiline Page 7 of 52 Medical Solutions 08.05 CS PS 24...
  • Page 8: Tolerance Data

    ± 8mm tolerance These tolerances apply to all dimensions indicated in these instructions, unless other tol- erances are specifically indicated after the value. The tolerance for torque values is ± 10%. LITHOSTAR Multiline RXL2-120.101.01.06.02 Siemens AG Page 8 of 52 08.05...
  • Page 9: Abbreviations

    The steps identified by these abbreviations are part of the maintenance report and should be checked off upon completion. NOTE Unless otherwise indicated, the maintenance interval is 12 months. Siemens AG RXL2-120.101.01.06.02 LITHOSTAR Multiline Page 9 of 52 Medical Solutions 08.05 CS PS 24...
  • Page 10: Inspection Of Exterior

    Image intensifier X-ray tube assembly cover Control PC Keyboard Mouse • Check the entire system for mechanical damage. • Replace or repair any defective parts. LITHOSTAR Multiline RXL2-120.101.01.06.02 Siemens AG Page 10 of 52 08.05 CS PS 24 Medical Solutions...
  • Page 11: Inspection Of Surroundings

    Ensure that there is free space around the unit to avoid collisions during unit movement. • Mark the locations of the additional devices or secure the units. Siemens AG RXL2-120.101.01.06.02 LITHOSTAR Multiline Page 11 of 52 Medical Solutions 08.05 CS PS 24...
  • Page 12 Safety inspections Safety inspections Remove the unit covers • Remove the side covers of the unit. LITHOSTAR Multiline RXL2-120.101.01.06.02 Siemens AG Page 12 of 52 08.05 CS PS 24 Medical Solutions...
  • Page 13: Electrical Safety

    The high voltage cables that can be touched and that conduct the tube current must be provided with a flexible, conductible shielding that has a maximum resistance of 1 ohm/m. Siemens AG RXL2-120.101.01.06.02 LITHOSTAR Multiline Page 13 of 52 Medical Solutions 08.05...
  • Page 14 • Switch off the signal generator. Do not reset the FL timer. Fluoro must switch off within 10 minutes. • Switch fluoroscopy OFF at the fluoro unit. LITHOSTAR Multiline RXL2-120.101.01.06.02 Siemens AG Page 14 of 52 08.05 CS PS 24...
  • Page 15 The protective conductor connection in the generator is the central protective conductor. (The protective conductor check must be performed again as a final step when the system is closed!) Siemens AG RXL2-120.101.01.06.02 LITHOSTAR Multiline Page 15 of 52 Medical Solutions 08.05 CS PS 24...
  • Page 16: Mechanical Safety

    • Check the limited range by moving unit G1 in the table cutout. Unit grip strip • Ensure that the unit grip strip is secure. LITHOSTAR Multiline RXL2-120.101.01.06.02 Siemens AG Page 16 of 52 08.05 CS PS 24...
  • Page 17 The yellow safety button lights up. • Move in the opposite direction of the unit blockage: Operate the safety button and the button on the tableside control panel. Siemens AG RXL2-120.101.01.06.02 LITHOSTAR Multiline Page 17 of 52 Medical Solutions 08.05 CS PS 24...
  • Page 18 Remove any interruptions in the light barrier. • Activate the safety cover on the shock wave head (L+R). Unit movements should not be possible. Yellow danger buttons light up. LITHOSTAR Multiline RXL2-120.101.01.06.02 Siemens AG Page 18 of 52 08.05 CS PS 24...
  • Page 19: Periodic Preventive Replacement Of Parts

    Check the battery on board D1 (≥ 2.7 V) and replace it at least once every 36 months. • Document this in the operating data report/log book. Siemens AG RXL2-120.101.01.06.02 LITHOSTAR Multiline Page 19 of 52 Medical Solutions 08.05 CS PS 24...
  • Page 20: Inspection For Wear

    Check and document the pressure value. See Shock Wave Pressure and Position Check Service Instructions SPL2-120.074.01. • Check functioning of switch S28 "adjustment aid attached". (Blocking of the W-drive) LITHOSTAR Multiline RXL2-120.101.01.06.02 Siemens AG Page 20 of 52 08.05 CS PS 24...
  • Page 21: Respiratory Belt

    - Holder for cable deflector 31 57 364 • Check the contacts for excessive melting loss. • Reattach the spark quenching chamber. • Reattach the RC element. Siemens AG RXL2-120.101.01.06.02 LITHOSTAR Multiline Page 21 of 52 Medical Solutions 08.05 CS PS 24...
  • Page 22: Changing The Water In The Coupling Circuit

    • After emptying the water tank, fill the coupling circuit in accordance with the start-up in- structions. LITHOSTAR Multiline RXL2-120.101.01.06.02 Siemens AG Page 22 of 52 08.05...
  • Page 23: Changing The Water In The Cooling Circuit

    Remove the float from the holder in the cover. • Clean the sealing surfaces of the float holder and the vent hole (Fig. 1 / p. 23) with flow- ing water. Siemens AG RXL2-120.101.01.06.02 LITHOSTAR Multiline Page 23 of 52 Medical Solutions 08.05 CS PS 24...
  • Page 24 Fill the cooling system and set the flow according to the start-up instructions. • Replace the water filter, part no. 70 52 012 F 0428, if necessary. • Refill the water supply tank after the water has been changed. LITHOSTAR Multiline RXL2-120.101.01.06.02 Siemens AG Page 24 of 52 08.05...
  • Page 25: Lubricating And Oiling The Linear Guides Of The Drives

    • Remove excess lubricant from the bearing blocks and linear guides. Reattach the removed cover plate. Guide rail, W-drive (oil) Guide rails, orbital (oil) Siemens AG RXL2-120.101.01.06.02 LITHOSTAR Multiline Page 25 of 52 Medical Solutions 08.05 CS PS 24...
  • Page 26 • Remove old or excess lubricant. • After oiling the W and O-drives, move them to the end positions several times to oil the guide rails evenly. LITHOSTAR Multiline RXL2-120.101.01.06.02 Siemens AG Page 26 of 52 08.05 CS PS 24...
  • Page 27: Cooling Unit In The Service Cabinet

    Check the complete cooling unit for leaks. • Ensure that the intake and exhaust air slots at the cooling unit are dust-free. Clean them if necessary. Siemens AG RXL2-120.101.01.06.02 LITHOSTAR Multiline Page 27 of 52 Medical Solutions 08.05 CS PS 24...
  • Page 28: Cabinet Ventilation

    Check whether the intake air and ventilation shafts are unobstructed when the cabinet is closed in order to ensure a fresh-air supply. • Remove any objects placed on the top of the cabinet! LITHOSTAR Multiline RXL2-120.101.01.06.02 Siemens AG Page 28 of 52 08.05...
  • Page 29: Memoskop

    Cleaning the ventilation slots • Remove dust from the bottom of the ventilation slots. • Perform a functional check of the fan installed in MEMOSKOP. Siemens AG RXL2-120.101.01.06.02 LITHOSTAR Multiline Page 29 of 52 Medical Solutions 08.05 CS PS 24...
  • Page 30: X-Ray Tube Assembly Fan

    Function Cleaning the ventilation slots • Check the X-ray tube assembly fan for proper functioning. • Clean the ventilation slots at the X-ray tube assembly housing. LITHOSTAR Multiline RXL2-120.101.01.06.02 Siemens AG Page 30 of 52 08.05 CS PS 24 Medical Solutions...
  • Page 31: Respiratory Gate Control

    The boards cannot be adjusted due to the complicated measuring setup! • Perform a functional check in accordance with Start-up Instructions RXL2-120.815... • Ensure proper functioning of the respiratory belt at both connection points. Siemens AG RXL2-120.101.01.06.02 LITHOSTAR Multiline Page 31 of 52 Medical Solutions 08.05 CS PS 24...
  • Page 32: Ecg Control

    The signal is displayed on the physiological monitor. A trigger is only re- leased when a QRS complex is detected! • Perform a functional check in accordance with Start-up Instructions RXL2-120.815... (See the Triggering Shock Waves section). LITHOSTAR Multiline RXL2-120.101.01.06.02 Siemens AG Page 32 of 52 08.05...
  • Page 33: Dose Control

    Operating value inspection Dose control Adjustment values • Check the dose during fluoroscopy. • (See Start-up Instructions RXL2-120.815... for the assigned value see the test report). Siemens AG RXL2-120.101.01.06.02 LITHOSTAR Multiline Page 33 of 52 Medical Solutions 08.05 CS PS 24...
  • Page 34: Generator

    - An initial brief spiked overshoot (e-function < 3 ms, depending on the length of the HV cable) is the charging current of the HV capacitors and can be ignored (does not flow over the tube and therefore does not affect dose). LITHOSTAR Multiline RXL2-120.101.01.06.02 Siemens AG Page 34 of 52 08.05...
  • Page 35: Pc Keyboard

    Functional inspections Functional inspections PC keyboard Key functions • Ensure proper functioning of all control keys on the keyboard in accordance with the op- erating instructions. Siemens AG RXL2-120.101.01.06.02 LITHOSTAR Multiline Page 35 of 52 Medical Solutions 08.05 CS PS 24...
  • Page 36: Triggering Shock Waves

    Functional inspections Triggering shock waves Triggering • Check shock wave triggering with and without respiratory and ECG triggering. (See the start-up instructions). LITHOSTAR Multiline RXL2-120.101.01.06.02 Siemens AG Page 36 of 52 08.05 CS PS 24 Medical Solutions...
  • Page 37: Footswitches

    (Electrical Safety / p. 13) section: Time - radiation indicator. • Check the footswitch at the unit and control console. (See the radiation indicator function). Siemens AG RXL2-120.101.01.06.02 LITHOSTAR Multiline Page 37 of 52 Medical Solutions 08.05 CS PS 24...
  • Page 38: Unit Footswitches - Urology

    Unit footswitches - urology Function • Check all footswitch functions in accordance with the operating instructions. • Check the cables and plug connections for damage. LITHOSTAR Multiline RXL2-120.101.01.06.02 Siemens AG Page 38 of 52 08.05 CS PS 24 Medical Solutions...
  • Page 39: Surveillance Monitor

    Basic settings • Check the basic settings and functioning of the physiological monitor. Take the appro- priate amplification into account when displaying the monitoring signals. Siemens AG RXL2-120.101.01.06.02 LITHOSTAR Multiline Page 39 of 52 Medical Solutions 08.05 CS PS 24...
  • Page 40: Tableside Control Panel

    Check the cables and plug connections for damage. • Ensure proper functioning of the tableside control panel of the control console. • Ensure proper functioning of the ultrasound tableside control panel. LITHOSTAR Multiline RXL2-120.101.01.06.02 Siemens AG Page 40 of 52 08.05...
  • Page 41: Control Pc Mouse

    • Ensure proper movement of the mouse pointer on the control monitor. • If necessary, remove the ball from the mouse and clean the sensor. Siemens AG RXL2-120.101.01.06.02 LITHOSTAR Multiline Page 41 of 52 Medical Solutions 08.05 CS PS 24...
  • Page 42: Component Inspection

    Component inspection 9.1.1 Television chain Image quality • Use the image quality test to check the image quality of the television chain. (See Start-up Instructions RXL2-120.815...). LITHOSTAR Multiline RXL2-120.101.01.06.02 Siemens AG Page 42 of 52 08.05 CS PS 24 Medical Solutions...
  • Page 43: Final Result Quality Inspection

    Save the current system parameters and delete fault storage. (See Operating the Service Software "Backup"). • Document the various counter values in the "Operational certificate" log book. Siemens AG RXL2-120.101.01.06.02 LITHOSTAR Multiline Page 43 of 52 Medical Solutions 08.05 CS PS 24...
  • Page 44: Collimator

    Check the predefined fixed collimator formats and adjust them if necessary in accor- dance with the setting instructions. • Check the settings of the collimator iris and adjust them if necessary in accordance with the setting instructions. LITHOSTAR Multiline RXL2-120.101.01.06.02 Siemens AG Page 44 of 52 08.05...
  • Page 45: Cassette Holder

    Check the centering of the cassette in the cassette holder with respect to the radiation field. • If necessary, adjust it in accordance with the setting instructions. Siemens AG RXL2-120.101.01.06.02 LITHOSTAR Multiline Page 45 of 52 Medical Solutions 08.05 CS PS 24...
  • Page 46: Isocenter

    Check the adjustment of the shock wave head. (See "Adjusting the isocenter" in the adjustment instructions). • If the ultrasound option is available, also check the isocenter with ultrasound. (See in- structions RXL-120.091..). LITHOSTAR Multiline RXL2-120.101.01.06.02 Siemens AG Page 46 of 52 08.05...
  • Page 47: Unit And System Cabinets

    Repair paint damage as necessary, for the paint number see Planning Guide RXL2-120.021.01. • Clean the tube cone from the inside (Only perform if the ultrasound option is available). Siemens AG RXL2-120.101.01.06.02 LITHOSTAR Multiline Page 47 of 52 Medical Solutions 08.05 CS PS 24...
  • Page 48: Log Book

    Log book Completeness • Check and document completed updates. • Fill out the system papers, check them for completeness, and add any missing informa- tion. LITHOSTAR Multiline RXL2-120.101.01.06.02 Siemens AG Page 48 of 52 08.05 CS PS 24 Medical Solutions...
  • Page 49: Covers

    Final Work Steps Final Work Steps 12.1 Covers Installing unit covers • Reattach all covers and check them for damage. Siemens AG RXL2-120.101.01.06.02 LITHOSTAR Multiline Page 49 of 52 Medical Solutions 08.05 CS PS 24...
  • Page 50: Unit Function

    - Replace them in the case of damage (see the spare parts list). Protective conductor test • Check the protective conductor when the system is closed in accordance with ARTD-002.731.17. (Also see the start-up instructions). LITHOSTAR Multiline RXL2-120.101.01.06.02 Siemens AG Page 50 of 52 08.05 CS PS 24...
  • Page 51 Changes to Previous Version Changes to Previous Version Toothed belt, cone lenght (CL) drive added to "Inspection for Wear." Siemens AG RXL2-120.101.01.06.02 LITHOSTAR Multiline Page 51 of 52 Medical Solutions 08.05 CS PS 24...
  • Page 52 Changes to Previous Version LITHOSTAR Multiline RXL2-120.101.01.06.02 Siemens AG Page 52 of 52 08.05 CS PS 24 Medical Solutions...

Table of Contents