Page 2
Assemblers and other persons who are not employed by or otherwise directly affiliated with or authorized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or one of its affiliates before attempting installation or service pro- cedures.
Page 3
Repair instructions ultrasound localizer RXL2-120.091..• Supplement for operating instructions Revision 06 is installed in systems < 07136 (UI SP 013 / 01). NOTE Siemens AG RXL2-120.033.01 Page 3 of 14 LITHOSTAR Multiline Medical Solutions Rev. 03 11.01 CS PS 24 System...
Page 4
The SONOLINE Prima does not require any special installation instructions. All you need to do is unpack the unit as described in the "unpacking procedure" affixed to the outside of the unit and attah the accessories. Mechanical modifications on LITHOSTAR Multiline Preparation •...
Page 5
2 screws M4 x 12,2 washers, and 2 lock nuts. Tab. 1 CAUTION When moving the W drive back, be careful not to damage the cables. Siemens AG RXL2-120.033.01 Page 5 of 14 LITHOSTAR Multiline Medical Solutions Rev. 03 11.01 CS PS 24 System...
Page 6
6 mm hole. Place any excessive cable length inside the holder. • • Remove any drill chips. Mount the probe holder. Tab. 2 LITHOSTAR Multiline RXL2-120.033.01 Page 6 of 14 Siemens AG System Rev. 03 11.01 CS PS 24 Medical Solutions...
Page 7
Screw the cable ties to the support bracket (Fig. 7) using washers and lock nuts. • Use cable ties to secure the cable (Fig. 7). Siemens AG RXL2-120.033.01 Page 7 of 14 LITHOSTAR Multiline Medical Solutions Rev. 03 11.01 CS PS 24 System...
Page 8
• Route the video cable and the fiber optic cable to the control cabinet. • Remove the EMC cover from N11 in front of boards D7 d/c, if existing. LITHOSTAR Multiline RXL2-120.033.01 Page 8 of 14 Siemens AG System Rev.
Page 9
Connect the corrugated tubing to the SONOLINE, if required. If the corrugated tubing has a plug, a wall support bracket is required. Siemens AG RXL2-120.033.01 Page 9 of 14 LITHOSTAR Multiline Medical Solutions Rev. 03 11.01 CS PS 24 System...
To set the date, time and name of the clinic, refer to the SONOLINE Adara Operating Instructions, Chapiter 4 System setup "General system settings". • Complete the operating instructions for the LITHOSTAR Multiline. • File the technical documents supplied in the system binder.
Page 11
Installing the Ultrasound Localizer Completion Protocol The ultrasound localizer option has been installed. Part Number: 31 27 011 Model No: LITHOSTAR Multiline Serial Number: ........ SKZ: ........ Customer: ........ Site: ........ Customer No.: ........ Name ( CSE ): ........ Telephone: ........
Page 12
Installing the Ultrasound Localizer Connect service PC to LITHOSTAR Multiline (also refer to RXL2-120.113.01...) Select Technical Info ... in menu Info Click Get from Unit Click Edit Enter the following information: TI/Speed Info: RXL2-120.033.01 Title: Option Ultrasound Date: When UI was performed...
Page 13
Paragraph ’Unpacking / assembling the ultrasound system’ - new paragraph inserted Page 10 Paragraph ’System start’ - sentence inserted Page 11 Old page deleted and new page inserted Siemens AG RXL2-120.033.01 Page 13 of 14 LITHOSTAR Multiline Medical Solutions Rev. 03 11.01 CS PS 24 System...
Page 14
Installing the Ultrasound Localizer This page intentionally left blank. LITHOSTAR Multiline RXL2-120.033.01 Page 14 of 14 Siemens AG System Rev. 03 11.01 CS PS 24 Medical Solutions...