Assemblers and other persons who are not employed by or otherwise directly affiliated with or autho- rized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or one of its affiliates before attempting installation or service procedures.
26 Speed not reached or exceeded 27 Motor moves in wrong direction 28 Motor actuated unsuccessfully three times 29 Faulty initialization of the ASIC Siemens AG RXL2-120.072.03 Page 1 of 2 LITHOSTAR Multiline Medical Solutions Rev. 04 06.02 CS PS 24 System...
49 Faulty initialization of the ASIC Board errors, general Error number: ↓ Meaning: 60 No first contact with HOST 61 TIMEOUT, data exchange disrupted 62 CHECKSUM_error LITHOSTAR Multiline RXL2-120.072.03 Page 2 of 2 Siemens AG System Rev. 04 06.02 CS PS 24...
26 Speed not reached or exceeded 27 Motor moves in wrong direction 28 Motor actuated unsuccessfully three times 29 Faulty initialization of the ASIC Siemens AG RXL2-120.072.03 Page 1 of 4 LITHOSTAR Multiline Medical Solutions Rev. 04 06.02 CS PS 24 System...
55 Motor does not move when actuated 56 Speed not reached or exceeded 57 Motor moves in wrong direction 58 Motor unsuccessfully actuated three times 59 Faulty initialization of the ASIC LITHOSTAR Multiline RXL2-120.072.03 Page 2 of 4 Siemens AG System Rev.
2 - 3 Board errors, general Error number: ↓ Meaning: 60 No first contact with HOST 61 TIMEOUT, data exchange disrupted 62 CHECKSUM_error Siemens AG RXL2-120.072.03 Page 3 of 4 LITHOSTAR Multiline Medical Solutions Rev. 04 06.02 CS PS 24 System...
Page 10
2 - 4 Motor controller D4b This page intentionally left blank. LITHOSTAR Multiline RXL2-120.072.03 Page 4 of 4 Siemens AG System Rev. 04 06.02 CS PS 24 Medical Solutions...
Data reception error from slave Serial transmitter not free DPR reception area error Timeout slave no. x Data reception error slave no. x Siemens AG RXL2-120.072.03 Page 1 of 4 LITHOSTAR Multiline Medical Solutions Rev. 04 06.02 CS PS 24 System...
Timeout error from the brightness value actual value signal device Data reception error of brightness actual value Serial transmitter not free Error in the DPR reception area LITHOSTAR Multiline RXL2-120.072.03 Page 2 of 4 Siemens AG System Rev. 04 06.02...
Error in block reception from ultrasound system Error in block transmission to ultrasound system Transmission buffer not empty (component) Error in the DPR reception area Siemens AG RXL2-120.072.03 Page 3 of 4 LITHOSTAR Multiline Medical Solutions Rev. 04 06.02 CS PS 24 System...
Page 14
3 - 4 Communication controller D6 This page intentionally left blank. LITHOSTAR Multiline RXL2-120.072.03 Page 4 of 4 Siemens AG System Rev. 04 06.02 CS PS 24 Medical Solutions...
Shock wave head, orbital X = am5 I.I. Image intensifier lift C arm Unit, longitudinal Unit, lateral ZR, ZL Table, up Shockwave head Siemens AG RXL2-120.072.03 Page 1 of 28 LITHOSTAR Multiline Medical Solutions Rev. 04 06.02 CS PS 24 System...
(switching the system ON/OFF). W = Warning System operation may continue without restrictions. I = Information These error numbers are for information only. They are not relevant to service. LITHOSTAR Multiline RXL2-120.072.03 Page 2 of 28 Siemens AG System Rev. 04 06.02...
ADC value out of tolerance / channel 13 008E ADC value out of tolerance / channel 14 008F ADC value out of tolerance / channel 15 Siemens AG RXL2-120.072.03 Page 3 of 28 LITHOSTAR Multiline Medical Solutions Rev. 04 06.02 CS PS 24 System...
Page 18
If, however, the software is present, then parameters are missing or the ver- sion numbers of the HOST software or the parameters in the EEPROM do not match. The error will be displayed after the download routine. Repeat the download routine. LITHOSTAR Multiline RXL2-120.072.03 Page 4 of 28 Siemens AG System Rev.
Error: Switch S2.8 After three unsuccessful attempts to transfer, (D1) = On transmission between the HOST and the ser- vice PC is terminated. Siemens AG RXL2-120.072.03 Page 5 of 28 LITHOSTAR Multiline Medical Solutions Rev. 04 06.02 CS PS 24 System...
NBF: Actuated or defective key NBF:Actuated or defective key 3281 NBF: Watchdog hit 3701 NBF Timeout 3702 Reception error of D6 from NBF LITHOSTAR Multiline RXL2-120.072.03 Page 6 of 28 Siemens AG System Rev. 04 06.02 CS PS 24 Medical Solutions...
Page 21
NBFUS: Timeout to board D6 3952 NBFUS: Keyboard buffer overflow 3960 NBFUS: actuated or defective key NBFUS: Actuated or defective 3981 NBFUS: Watchdog hit Siemens AG RXL2-120.072.03 Page 7 of 28 LITHOSTAR Multiline Medical Solutions Rev. 04 06.02 CS PS 24 System...
2 4142 Control error / W am 3 4143 Control error / O am 4 4145 Control error / X am 5 LITHOSTAR Multiline RXL2-120.072.03 Page 8 of 28 Siemens AG System Rev. 04 06.02 CS PS 24 Medical Solutions...
Page 23
9 417A Invalid data for motor initialization / SWO am 10 417B Invalid data for motor initialization / I.I. am 11 Siemens AG RXL2-120.072.03 Page 9 of 28 LITHOSTAR Multiline Medical Solutions Rev. 04 06.02 CS PS 24 System...
Page 24
Motor rotation in reverse direction / W am 3 41B3 Motor in reverse direction / O am 4 41B5 Motor in reverse direction / X am 5 LITHOSTAR Multiline RXL2-120.072.03 Page 10 of 28 Siemens AG System Rev. 04 06.02 CS PS 24...
Page 25
ADC/MC pos. comparison failure / TR am 9 428A ADC/MC pos. comparison failure / SWO am 10 428B ADC/MC pos. comparison failure / I.I. am 11 Siemens AG RXL2-120.072.03 Page 11 of 28 LITHOSTAR Multiline Medical Solutions Rev. 04 06.02 CS PS 24 System...
Page 26
Error in averaging / 42AA Error in averaging / SWO Error in averaging / 42AB Error in averaging / I.I. Error in averaging / I.I. LITHOSTAR Multiline RXL2-120.072.03 Page 12 of 28 Siemens AG System Rev. 04 06.02 CS PS 24...
< kA D19/X1,b16 = high and yellow LED D19/V32 lights up actual 5283 D19 U is too high D19/X1,b12 = high and red LED D19/V9 lights up residual LITHOSTAR Multiline RXL2-120.072.03 Page 14 of 28 Siemens AG System Rev. 04 06.02...
Page 29
< > kV actual selected 5383 D19 kA > kA actual 5391 D91 Error memory see chapter 8 5392 D19 Error memory Siemens AG RXL2-120.072.03 Page 15 of 28 LITHOSTAR Multiline Medical Solutions Rev. 04 06.02 CS PS 24 System...
Start up timeout / Image Store 6136 Start up timeout / Shock Wave System 6137 Start up timeout / Actual Brightness Value 6138 Start up timeout / Ultrasound LITHOSTAR Multiline RXL2-120.072.03 Page 16 of 28 Siemens AG System Rev. 04 06.02...
Invalid value for tube loading 911B Invalid value for acknowledgement byte 9120 Invalid obj.ID for I.I. functional unit 9121 Invalid value for I.I. func. unit: ZOOM LITHOSTAR Multiline RXL2-120.072.03 Page 18 of 28 Siemens AG System Rev. 04 06.02 CS PS 24...
Replace board D2 in the Memoskop. Replace the backplane board D5. Replace board D1 in the Memoskop (Observe the configuration 50/60 Hz, 3 or 100 images). Siemens AG RXL2-120.072.03 Page 19 of 28 LITHOSTAR Multiline Medical Solutions Rev. 04 06.02 CS PS 24 System...
Page 34
Keyboard interface test, feedback error Replace the keyboard or the keyboard cable of the Memoskop. Replace board D1 in the Memoskop (Observe the configuration 50/60 Hz, 3 or 100 images). LITHOSTAR Multiline RXL2-120.072.03 Page 20 of 28 Siemens AG System Rev.
Page 35
Correct connections transformer T1 - plug D1/X13. Replace board D1 in the Memoskop (Observe the configuration 50/60 Hz, 3 or 100 images). Siemens AG RXL2-120.072.03 Page 21 of 28 LITHOSTAR Multiline Medical Solutions Rev. 04 06.02 CS PS 24 System...
Page 36
(Observe the configuration 50/60 Hz, 3 or 100 images). E812 Memoskop voltage test: Voltage VPPmax. too high. No test. Replace board D1 in the Memoskop (Observe the configuration 50/60 Hz, 3 or 100 images). LITHOSTAR Multiline RXL2-120.072.03 Page 22 of 28 Siemens AG System Rev.
Page 37
Check the LED displays on the Memoskop. Check the connection: Service PC HOST/D1. Try download routine again. Adjust the service PC, if necessary Image memory (errors from Multiline) Siemens AG RXL2-120.072.03 Page 23 of 28 LITHOSTAR Multiline Medical Solutions Rev. 04 06.02 CS PS 24 System...
Page 38
Memoskop Board D2 Help: FEPROM defective + 15 V supply voltage, X13 Board D2, D1 defective ED36 Memoskop Board D2 FEPROMS all bytes already 0 LITHOSTAR Multiline RXL2-120.072.03 Page 24 of 28 Siemens AG System Rev. 04 06.02...
Page 39
Memoskop Board D2 FEPROM checksum error during download Checksum of FEPROMs incorrect End of download of Memoskop Board D2 Help: FEPROM defective Siemens AG RXL2-120.072.03 Page 25 of 28 LITHOSTAR Multiline Medical Solutions Rev. 04 06.02 CS PS 24 System...
Acquisition active without fluoro EF70 Acquisition timeout during fluoro EF80 Timeout acknowledgement "SH_Ready" (SH = from Slave to D1) EF90 Timeout acknowledgement "SH_Stop" LITHOSTAR Multiline RXL2-120.072.03 Page 26 of 28 Siemens AG System Rev. 04 06.02 CS PS 24 Medical Solutions...
The system is not initialized F812 Error in writing process F813 Error in comparison of desired and actual values F814 Error while writing checksum Siemens AG RXL2-120.072.03 Page 27 of 28 LITHOSTAR Multiline Medical Solutions Rev. 04 06.02 CS PS 24 System...
FE92 Parameter error "Direct_mode" FE93 Inadmissible value in "FE_BV_ZOOM" Unacceptable value in "FE_BV_Zoom" FE94 Inadmissible value in "XCS_STAT()" Unacceptable value in "XCS_STAT()" LITHOSTAR Multiline RXL2-120.072.03 Page 28 of 28 Siemens AG System Rev. 04 06.02 CS PS 24 Medical Solutions...
"Interface-Error: SEND-BUFFER FULL" Meaning: The operating PC has generated more data blocks than can be sent and thus triggered an internal data overflow; Siemens AG RXL2-120.072.03 Page 1 of 2 LITHOSTAR Multiline Medical Solutions Rev. 04 06.02 CS PS 24 System...
POLYDOROS LX 30/50 Troubleshooting instructions R63-020.072... POLYDOROS SX 65/80 Troubleshooting instructions R63-055.072... and the VIDEOMED DI/DIM Service instructions RX52-022.061... contain information on the causes of errors. LITHOSTAR Multiline RXL2-120.072.03 Page 2 of 2 Siemens AG System Rev. 04 06.02 CS PS 24...
Page 45
HOST system Decimal point changes frequently: ACS net is not stable; along with the message number 30 Decimal point flashes: correct program run Siemens AG RXL2-120.072.03 Page 1 of 2 LITHOSTAR Multiline Medical Solutions Rev. 04 06.02 CS PS 24 System...
Page 46
6 - 2 XCS controller D2 This page intentionally left blank. LITHOSTAR Multiline RXL2-120.072.03 Page 2 of 2 Siemens AG System Rev. 04 06.02 CS PS 24 Medical Solutions...
Spark gap ignition occurred twice without triggering. For settings on board D19, refer to instructions "Repair of shock NOTICE wave system" RXL2-120.091... Siemens AG RXL2-120.072.03 Page 1 of 8 LITHOSTAR Multiline Medical Solutions Rev. 04 06.02 CS PS 24 System...
Page 48
Charging unit Unplug high-voltage cable K26 not charging up charging up High-voltage cable K26 or Charging unit shock wave generator defective defective LITHOSTAR Multiline RXL2-120.072.03 Page 2 of 8 Siemens AG System Rev. 04 06.02 CS PS 24 Medical Solutions...
Replace spark gap module Yellow LED V33 (spontaneous ignition of the spark gap) Board D1 Display: 5281 Reset on keyboard with ESC Error occurs repeatedly Replace spark gap module LITHOSTAR Multiline RXL2-120.072.03 Page 4 of 8 Siemens AG System Rev. 04 06.02...
Error not removed Check adjustment of the circuit board D19 Error not removed Replace board D19 Siemens AG RXL2-120.072.03 Page 5 of 8 LITHOSTAR Multiline Medical Solutions Rev. 04 06.02 CS PS 24 System...
Page 52
17.1 17.4 17.8 18.1 3.64 V ± 0.03 V 18,3 10.0 18.4 10.5 18.7 11.0 18.9 11.5 19.3 12.0 19.6 Tab. 1 LITHOSTAR Multiline RXL2-120.072.03 Page 6 of 8 Siemens AG System Rev. 04 06.02 CS PS 24 Medical Solutions...
Page 53
7 - 7 K108. N12 K108. charging unit Energy levels 10.0 10.5 11.0 11.5 12.0 Tab. 2 K108 pin assignment on the charging unit Siemens AG RXL2-120.072.03 Page 7 of 8 LITHOSTAR Multiline Medical Solutions Rev. 04 06.02 CS PS 24 System...
Page 54
7 - 8 Error displays on D19 This page intentionally left blank. LITHOSTAR Multiline RXL2-120.072.03 Page 8 of 8 Siemens AG System Rev. 04 06.02 CS PS 24 Medical Solutions...
This makes it likely that board D91 has responded. For settings on board D91, refer to the instructions "Maintenance NOTICE of shockwave system" RXL2-120.091... Siemens AG RXL2-120.072.03 Page 1 of 4 LITHOSTAR Multiline Medical Solutions Rev. 04 06.02 CS PS 24 System...
Page 56
X5 and X5 < kV > kV act/2 act/2 Check leads > kV < kV act/2 act/2 Replace shock wave generator LITHOSTAR Multiline RXL2-120.072.03 Page 2 of 4 Siemens AG System Rev. 04 06.02 CS PS 24 Medical Solutions...
Charging unit not charging up Replace shock wave generator Charging unit charging up High voltage cable K26 or Charging unit Shock wave generator defective defective Siemens AG RXL2-120.072.03 Page 3 of 4 LITHOSTAR Multiline Medical Solutions Rev. 04 06.02 CS PS 24 System...
Page 58
8 - 4 Error displays on D91 This page intentionally left blank. LITHOSTAR Multiline RXL2-120.072.03 Page 4 of 4 Siemens AG System Rev. 04 06.02 CS PS 24 Medical Solutions...
Page 59
Errors without error numbers 9 - 1 Letters appear scattered across the monitor. Cause: defective text module on board D1 in the Memoskop. Siemens AG RXL2-120.072.03 Page 1 of 2 LITHOSTAR Multiline Medical Solutions Rev. 04 06.02 CS PS 24 System...
Page 60
9 - 2 Errors without error numbers This page intentionally left blank. LITHOSTAR Multiline RXL2-120.072.03 Page 2 of 2 Siemens AG System Rev. 04 06.02 CS PS 24 Medical Solutions...
Page 61
Revision level increased, table of contents adjusted Chapter 4 - 27 Error F812 - new inserted Error F813 - new inserted Error F814 - new inserted Siemens AG RXL2-120.072.03 Page 1 of 2 LITHOSTAR Multiline Medical Solutions Rev. 04 06.02 CS PS 24 System...
Page 62
10 - 2 Changes as compared to previous version This page intentionally left blank. LITHOSTAR Multiline RXL2-120.072.03 Page 2 of 2 Siemens AG System Rev. 04 06.02 CS PS 24 Medical Solutions...