GE MEDICAL SYSTEMS
REV 5
1 6
Field Light Switch Replacement
ILLUSTRATION 1-9
BRAKE SWITCH REPLACEMENT
1. Place horizontal arm in lock position.
2. Shut power off.
3. Use a small screwdriver and very carefully pry the field light switch out of the col
limator control panel. The switch press fits into the panel and will easily snap out
of the panel. A short length of two wires will be attached to the switch. Use care
not to scratch the panel during this procedure. See Illustration 1-2.
4. Unsolder the two wires.
5. Resolder two wires to the terminals on new replacement switch.
6. Carefully insert the switch and wires into the panel and press into place.
SWITCH RETAINING
PLATE
BUTTON RETURN
SPRING
MICRO
SWITCH
1-12
(MODEL 2169360, 2236420 & 2275938 SERIES)
DIRECTION 2173225-100
(2) PHILLIPS SCREWS
AMX-4+ SERVICE
SWITCH
BUTTON