10 3 4 Stator Cable

ILLUSTRATION 10-13
STATOR CABLE ROUTE
TO FIRST TY-RAP
ILLUSTRATION 10-14
STATOR CABLE ROUTE
ON ANODE CABLE
1. The stator cable starts at the tube, passes under the cathode cable and joins the
anode cable just before clamp L." There is a large loop of cable before reaching
clamp point L." The cables are ty-rapped on each side of clamp point L " as
shown in Illustration 10-13.
CATHODE
STATOR CABLE
CLAMP POINT L "
2. The stator cable follows the anode cable from clamp point L " to strap point M"
with two ty-raps in between. See Illustration 10-14.
STRAP POINT G
(MODEL 2169360, 2236420 & 2275938 SERIES)
CLAMP POINT H
10-11
AMX-4+ SERVICE
DIRECTION 2173225-100
ANODE
CATHODE
COLLIMATOR CABLE
STRAP POINT M