GE AMX-4+ Technical Manual page 111

Hide thumbs Also See for AMX-4+:
Table of Contents

Advertisement

GE MEDICAL SYSTEMS
REV 5
8 4
Handle Drive Enable
Switch Installation
Note:
This procedure covers installation of either the right or left drive enable
switch in the handle assembly.
1. See Illustration 8-3, Detail B",for the drive enable microswitch.
2. Place the drive enable microswitch on its bracket and secure it in place with two
screws.
3. Feed drive enable switch leads through back of end casting to connector.
4. Slide end cap along handle into position on end casting.
5. Secure end cap to end casting with four hex socket button head capscrews.
6. Connect drive enable switch leads (blue, yellow or gray) to connector (positions
1 & 2 in connector). See Illustration 8-2.
7. Secure enable switch leads and other leads to end casting support bracket with
a ty-rap. See Illustration 8-1.
8. Close top cover. Refer to Section 5-6.
8-5
(MODEL 2169360, 2236420 & 2275938 SERIES)
DIRECTION 2173225-100
AMX-4+ SERVICE