GE MEDICAL SYSTEMS
REV 5
SECTION 10
SERVICING THE POWER CORD AND REEL ASSEMBLY AND CABLE ROUTING
10 1
Cord/Cord Reel Removal
10 1 1 Cord Reel Removal (Domestic Models)
ILLUSTRATION 10-1
CORD REEL REMOVAL/INSTALLATION
HEX WASHER
HEAD SCREW
This section covers the removal and installation of the main power cord reel, and the
horizontal arm/vertical column cable routing.
1. Shut power off by turning circuit breaker OFF.
2. Open top cover. Refer to Section 5-5.
3. Remove four hex nuts attaching low voltage power supply to mainframe and slide
it backward out of mainframe housing. See Illustration 10-1.
4. Disconnect cord reel leads from F1, F2 and ground stud.
5. Remove two internal hex cap screws retaining grommet in cut-out in mainframe
and remove cable and grommet.
6. Remove two hex washer head screws and lock nuts holding cord reel to side of
low voltage power supply housing.
7. Lift off cord reel pulling leads through grommet in low voltage power supply
housing.
F1
FUSE
HOLDERS
BALL
GROMMET
CORD REEL
10-1
(MODEL 2169360, 2236420 & 2275938 SERIES)
DIRECTION 2173225-100
TERMINALS
GROMMET IN
LOW
VOLTAGE
POWER
SUPPLY
HOUSING
F2
MAINFRAME
AMX-4+ SERVICE
GROUND