GE MEDICAL SYSTEMS
REV 5
ILLUSTRATION 5-5
LATCH WIRING
CAUTION
4. Disconnect leads from terminal strip (see Illustration 5-5):
D
Sensor Board: three leads to quick disconnect connectors. (#1, #2 and #3).
D
Solenoid: two leads to terminal strip. Tag and label terminals. Must recon
nect properly on #3 & #4 or will blow diode in solenoid.
WHITE
5. Remove latch assembly from mainframe by removing four hex socket head
capscrews.
6. Remove latch block from spacer by removing four hex socket head capscrews.
7. Push spring pin out of latch block.
8. Remove latch.
9. Remove latch spring by removing binding head screw.
10. Remove hex nut from top of solenoid.
11. Remove solenoid from recess in latch block.
12. Remove the two flat head screws in side of latch base.
Use care to avoid electrostatic discharge damage to the Hall effect sensor cir
cuit board.
13. Remove the cover and the Hall effect sensor circuit board.
(MODEL 2169360, 2236420 & 2275938 SERIES)
5-6
AMX-4+ SERVICE
DIRECTION 2173225-100
TERMINAL STRIP